Chapter 4 Inspection
35
HIGH FLOW INSUFFLATION UNIT UHI-3
Chapter 4 Inspection
• Before each case, inspect this instrument as instructed
below. Inspect other equipment to be used with this
instrument as instructed in their respective instruction
manuals. Should the slightest irregularity be suspected, do
not use the instrument and see Chapter 7, “Troubleshooting”.
If the irregularity is still suspected after consulting Chapter 7,
contact Olympus.
Damage or irregularity may compromise patient or user
safety and may result in more severe equipment damage.
• This instrument was not disinfected or sterilized before
shipment. Before using this instrument for the first time,
reprocess it according to the instructions given in Chapter 6,
“Care, Storage and Disposal”.
4.1
Inspection of the UHI-3 main unit
Power supply
1.
Press the POWER switch to turn the unit ON. Confirm that all indicator
lamps light and three tones sound.
2.
Confirm that the stop lamp lights up. Confirm that the last abdominal
pressure setting is recalled, the insufflation mode is set to LOW, and the
Volume Indicator shows “0.0 L”.
If the above indications fail, the equipment may be faulty.
Immediately turn the POWER switch OFF, disconnect the
power cord from the hospital grade receptacle (wall mains
outlet) and contact Olympus.