Chapter 3 Installation and Connection
33
HIGH FLOW INSUFFLATION UNIT UHI-3
3.7
Connecting the Electrosurgical Unit (UES-40) or
the SonoSurg generator (SonoSurg-G2)
This product, when combined with the UES-40 or the SonoSurg-G2 via aeration
cable MAJ-877, can exhaust fumes and mist generated in the body cavity by
synchronizing the ultrasonic or high frequency output and the exhaust function.
• When using the automatic exhaust, also refer to the
instruction manual of the UES-40 or the SonoSurg-G2.
Incorrect usage or equipment connection may compromise
the functionality and performance of this feature and could
lead to equipment damage and/or patient injury.
• Refer to the instruction manuals for the SonoSurg-G2, the
UES-40 and the instruments to be used with them
(endoscope, light source, ancillary equipment, etc.).
Attach the connectors of the aeration cable (MAJ-877) to the system connector
of the UHI-3 and the system connector of the UES-40 or the system-A connector
of the SonoSurg-G2. Then tighten the fixing screws on both connectors (see
Figure 3.12).