3.7 Checking the MRC level and entering the check result
71
OER-Elite OPERATION MANUAL
Ch.3
Checking the MRC level
1
Press the foot pedal to open the lid.
2
Uncap the disinfectant solution nozzle of the
reprocessing basin by turning the disinfectant
solution nozzle cap in the direction shown right.
Figure 3.45
3
Insert the test strip all the way into the
disinfectant solution nozzle and check the MRC
as described in the test strip instruction
manual. If the concentration is lower than the
effective level, replace the disinfectant solution
by following the instructions in Section 8.2,
“Replacing the disinfectant solution”.
Figure 3.46
NOTE
• Under high temperature and extreme temperature fluctuations, test strip may result
in fail. In such cases, Olympus recommends performing mix LCG and check the
MRC level again.
• Depending on conditions of internal valve, there is a possibility that no disinfectant
solution exist in the disinfectant solution nozzle. In the following cases, perform the
mix LCG, then check the MRC level again.
No reaction is observed on the chemical indicator after checking the MRC level.
Reaction is observed only from the edge parts of the chemical indicator after
checking the MRC level.
If the disinfectant solution in the disinfectant solution nozzle is found insufficient for
checking the MRC level on a regular basis, contact Olympus.
Disinfectant solution nozzle cap