4.12 LCG replacement indicator
97
OER-Elite OPERATION MANUAL
Ch.4
NOTE
For details of the indicator, refer to Section 3.2, “Alarm indicator”.
3
If “Edit” button is pressed in the previous screen, following screen is displayed. Edit
the number of days and cycles, and press “OK” button.
No.
Item/Button
Setting options
Description
1
Days
1 – 5 [days]
(default: 3)
Sets the set value of days for the LCG replacement indicator.
Pressing the “+” button increments the setting and pressing the
“–” button decrements it.
2
Cycles
1 – 35 [usages]
(default: 15)
Sets the set value of cycles for the LCG replacement indicator.
Pressing the “+” button increments the setting and pressing the
“–” button decrements it.
3
Cancel
–
Returns to the Setting menu without saving the edited information.
4
OK
–
Returns to the Setting menu after saving the edited information.
4
Press the “Save” button to return to the Setting
menu after saving the edited information.
Figure 4.2
1
3
4
2