Tier-1
Tier-2
Bottom
6
5
Two-tier multi-port design ensures input/output flexibility.
■
IX2 Two-tier optical path
Optical path as seen from
the left side of
the microscope
Objective
Optical path as seen from
the front of
the microscope
Mirror unit
Tube lens
Right side port
Left side port
Lower back port
Upper Tier Lightpath Selection (optional)
Located between objective and tube lenses so a parallel
pencil of rays can be obtained or introduced. Primary
image can be gained by adding a tube lens. Inserting
optical components such as a dichromatic mirror does
not produce a double image. (The alternative of the right
side port)
Lower Tier Lightpath Selection (included)
Located below the tube lens inside the frame, this tier
allows primary image access to either the left side port
or lower back port.
Bottom Lightpath Selection
A direct primary image can be obtained without any
reflections and any optical components.
Tier-1:
Upper optical path
selection
Tier-2:
Lower optical path
selection
Bottom optical
path selection
■
Left side port
This port offers a high quality
primary image, located in 102 mm
distance from the microscope
frame for maximum flexibility in
mounting filter wheels or any kinds
of camera adapters.
■
Dual port camera adapter /
U-DPCAD* (C-mount,
left side port)
This unique dual port adapter
enables the provision of two
primary images suitable for live cell
imaging.
* optional unit
The input/output of a parallel pencil of rays and the multiple port
structure for gaining the primary image are designed internally in the
form of tiers. To maximize the possible wavelength width, the
optical path branching of each tier is also compatible with the near-
infrared spectrum. Even when more than one port is being used
simultaneously, there is no change in the stage height; as a result,
rigidity and illumination performance remain constant.
Two-tier optical design is also near-IR compatible
Optical Port system
102 mm from left side port
mounting position to primary image.
■
Lower back port
■
Left side port
■
Bottom port
■
Right side port
■
Binocular port
102 mm
■
Lower back port
Enables the attachment of a
C mount camera. Only 1X
magnification is available.
■
Bottom port /
IX2-TVR (T-mount)
Primary image access is also
available at the microscope bottom.
IR compatible
IR compatible
IR compatible
IX2-SIDEPORT 現行vsエコ化AWマルチ 計算値 obなし
0
10
20
30
40
50
60
70
80
90
100
300
400
500
600
700
800
900
1000
波長(nm)
透過率%
IX2-SIDE現行-A
IX2-SIDEエコ化(KTL-ECO)-AW
0
10
20
30
40
50
60
70
80
90
100
300
400
500
600
700
800
900
1000
Wavelength (nm)
T
ransmittance (%)
Current IX2 left side port
New IX2 left side port
IX2 Side port: Transmittance improved by new coating
With the introduction of the new UIS2 optical system, the IX2 series
offers improved IR transmittance for the side port, back port and
bottom port, providing versatile, high-performance response to
future research demands.
Improved near-infrared transmission
■
Right side port / IX2-RSPC-2
The right side port unit (IX2-RSPC-2:
option, F.N.: 16) comes with a tube
lens and accepts a C-mount CCD
camera.