6

5

Two-tier multi-port design ensures input/output flexibility.

IX2 Two-tier optical path

Optical path as seen from

the left side of 

the microscope

Objective

Optical path as seen from

the front of 

the microscope

Mirror unit

Tube lens

Right side port

Left side port

Lower back port

Upper Tier Lightpath Selection (optional)

Located between objective and tube lenses so a parallel
pencil of rays can be obtained or introduced. Primary
image can be gained by adding a tube lens. Inserting
optical components such as a dichromatic mirror does
not produce a double image. (The alternative of the right
side port)

Lower Tier Lightpath Selection (included)

Located below the tube lens inside the frame, this tier
allows primary image access to either the left side port
or lower back port.

Bottom Lightpath Selection 

A direct primary image can be obtained without any
reflections and any optical components. 

Tier-1: 

Upper optical path

selection

Tier-2: 

Lower optical path

selection

Bottom optical

path selection

Left side port

This port offers a high quality
primary image, located in 102 mm
distance from the microscope
frame for maximum flexibility in
mounting filter wheels or any kinds
of camera adapters.

Dual port camera adapter /

U-DPCAD* (C-mount, 
left side port)

This unique dual port adapter
enables the provision of two
primary images suitable for live cell
imaging. 

* optional unit

The input/output of a parallel pencil of rays and the multiple port
structure for gaining the primary image are designed internally in the
form of tiers. To maximize the possible wavelength width, the
optical path branching of each tier is also compatible with the near-
infrared spectrum. Even when more than one port is being used
simultaneously, there is no change in the stage height; as a result,
rigidity and illumination performance remain constant.

Two-tier optical design is also near-IR compatible

Optical Port system

102 mm from left side port
mounting position to primary image. 

Lower back port

Left side port

Bottom port

Right side port

Binocular port

102 mm

Lower back port

Enables the attachment of a 
C mount camera. Only 1X
magnification is available.

Bottom port / 

IX2-TVR (T-mount)

Primary image access is also
available at the microscope bottom.

IR compatible

IR compatible

IR compatible

IX2-SIDEPORT 現行vsエコ化AWマルチ 計算値 obなし 

0

10

20

30

40

50

60

70

80

90

100

300

400

500

600

700

800

900

1000

波長(nm)

透過率%

IX2-SIDE現行-A

IX2-SIDEエコ化(KTL-ECO)-AW

0

10

20

30

40

50

60

70

80

90

100

300

400

500

600

700

800

900

1000

Wavelength (nm)

T

ransmittance (%)

Current IX2 left side port 

New IX2 left side port

IX2 Side port: Transmittance improved by new coating

With the introduction of the new UIS2 optical system, the IX2 series
offers improved IR transmittance for the side port, back port and
bottom port, providing versatile, high-performance response to
future research demands.

Improved near-infrared transmission

Right side port / IX2-RSPC-2

The right side port unit (IX2-RSPC-2:
option, F.N.: 16) comes with a tube
lens and accepts a C-mount CCD
camera.

Содержание IX71

Страница 1: ...Laboratory for genomic Reprogramming Head of Laboratory Riken Kobe Institute Center for Developmental Biology CDB P13 Yuji Abe M D Ph D The 1st Department of Obstetrics Gynecology School of Medicine...

Страница 2: ...wavelength range and flat high transmittance this new system sets a new world standard of fluorescence performance efficiently detecting even faint fluorescence signals without damaging the cell optim...

Страница 3: ...lar transmission in the near infrared range is significantly enhanced Overall performance all across the wavelength range is ideally suited for today s most demanding research applications High N A ob...

Страница 4: ...put output of a parallel pencil of rays and the multiple port structure for gaining the primary image are designed internally in the form of tiers To maximize the possible wavelength width the optical...

Страница 5: ...filter wheels on the excitation and emission sides However this kind of mirror encounters the problem that each fluorescence image gets darker as the number of color dyes increase because the transmi...

Страница 6: ...g adapter U DULHA Allows simultaneous attachment of two light sources such as halogen and mercury Selection mirror is replaceable for custom applications Fluorescence illumination light source Reflect...

Страница 7: ...DICT U DICTS Thin specimen Big shearing value Thick specimen Small shearing value 10X 40X 100X U DICTHC for superior contrast with thin specimen observation U DICTHR for superior resolution with thic...

Страница 8: ...s DIC optical elements for IX2 MLWCD and applicable objectives 1 Objective with compensation for 1 mm plastic dish plus 0 5 mm thick thermoplate 2 Objective with compensation ring for 0 2 mm thick cov...

Страница 9: ...ffers different magnification without switching the objective lens 1 6X is standard IX71 81 and 2 0X is optional In addition to maximizing rigidity of IX2 microscope frame Olympus simplified or shorte...

Страница 10: ...pen close and camera vs visual observation Parfocal compensation function among objectives This function allows the focus point to be matched from low to high magnification objectives Refocusing each...

Страница 11: ...e cell injection Built in stage warming plate Objective heater 5 CO2 supply tube with 4 outer diameter 2 inner diameter and 400 mm length Not available in some areas Thermoplate MATS series This therm...

Страница 12: ...llumination with twin scanners TIRFM Total Internal Reflection Fluorescence Microscopy Since 1997 Olympus has been a market leader in objective based Total Internal Reflection microscope that allows a...

Страница 13: ...system from an arc lamp source ARC EVA SYSTEM DIAGRAM Main specifications Microscope Research inverted system microscope IX71 Fluorescence Arc illumination total internal reflection illuminator fluor...

Страница 14: ...order Easy system upgrade by attaching double lamphouse illuminator IX2 RFAW to IX2 series inverted microscope Photoactivation illuminator for inverted microscopy The novel Kaede gene is useful in bi...

Страница 15: ...urret IX2 RFAC 6 positions in a rotating turret built in shutter Light source 100 W Hg lamp housing and transformer or 75 W Xe lamp housing and transformer IX71 specifications Microscope body Revolvin...

Страница 16: ...tage IX SUSP Stage plate Incubator IX2 KSP Narrow plain stage CK2 SS Side stage U SIP Split primary image camera port Cube cassette for full image split images U IFFH Focus handle Interface U ZPCB2 Z...

Отзывы: