44
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Flushing detergent solution into the elevator wire channel
1.
Connect the washing tube to the elevator channel plug (see Figure 3.31).
Figure 3.31
2.
Using the 5 cm
3
(5 ml) syringe, inject 15 cm
3
(15 ml) of detergent solution
into the elevator wire channel via the washing tube (see Figure 3.32).
Figure 3.32
3.
Disconnect the washing tube from the endoscope and immerse it in
detergent solution.
Elevator channel plug
Washing tube
5 cm3 (5 ml) syringe
Elevator wire channel
Endoscope connector
Washing tube
Elevator channel
plug