Chapter 2 Compatible Reprocessing Methods and Chemical Agents
11
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
ETO gas exposure parameters
Table 2.2
Gas mixture
20% ETO/80% CO
2
2.6
Steam sterilization (autoclaving) of accessories
The accessories listed as compatible with steam sterilization (autoclaving) in
Table 2.1 can be sterilized by steam within the parameters given in Table 2.3.
When steam sterilizing, follow the cleaning, disinfection and sterilization
protocols of your hospital and the sterilization equipment manufacturer’s
instructions.
The results of sterilization depend on various factors such as
how the sterilized instrument was packed or the positioning,
method of placing and loading of the instrument in the
sterilization device. Please verify the sterilization effects by
using biological or chemical indicators. Also follow the
guidelines for sterilization issued by medical administrative
authorities, public organizations or the infection management
sections at each medical facility, as well as the instruction
manual of the sterilization device.
• Do not steam sterilize the endoscope. Steam sterilization
(autoclaving) will severely damage the endoscope.
Process
Parameters
ETO gas sterilization
Temperature
57°C (135°F)
Pressure
0.1 – 0.17 MPa (1 – 1.7 kgf/cm
2
)
(16 – 24 psig)
Humidity
55%
Exposure time
1.75 hours
ETO gas
concentration
0.6 – 0.7 mg/cm
3
(600 – 700 mg/l)
Aeration
Minimum aeration
parameters
12 hours in an aeration chamber
at 50 – 57°C (122 – 135°F) or
7 days at room temperature