DMTA-10084-01EN, Rev. 2, November 2018
Pulser/Receiver Adjustments
59
6. Pulser/Receiver Adjustments
This chapter describes how to adjust the EPOCH 6LT pulser/receiver.
6.1
Adjusting the Gain (System Sensitivity)
To adjust the gain
1.
On the live A-scan screen, press the Tab key to select the
Gain
adjustment box
2.
Turn the adjustment knob to set the gain (system sensitivity).
The total gain (system sensitivity) is 110 dB.
Figure 6-1 Highlighted Gain adjustment box
NOTE
Содержание EPOCH 6LT
Страница 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Страница 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Страница 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Страница 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Страница 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Страница 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Страница 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Страница 109: ...DMTA 10084 01EN Rev 2 November 2018 Calibration 99 Figure 8 7 Calibration Successfully completed...
Страница 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Страница 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Страница 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Страница 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Страница 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Страница 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...