![Olympus EPOCH 6LT Скачать руководство пользователя страница 189](http://html2.mh-extra.com/html/olympus/epoch-6lt/epoch-6lt_user-interface-manual_3934183189.webp)
DMTA-10084-01EN, Rev. 2, November 2018
Software Features and Options
179
DGS/AVG curve gain adjustments are applied to both the echo height and curve
height to maintain the amplitude ratio and, therefore, the sizing comparisons.
Figure 11-23 on page 179 shows a DGS/AVG setup with curve gain in use to provide
accurate echo-amplitude measurement by placing the echo close to 80 % FSH.
Figure 11-23 Gain curve adjusted DGS
11.4.5
Registration Level Adjustment
The registration level of the DGS/AVG curve defines the height of the main curve. The
curve height represents the amplitude from a flat-bottom hole with a diameter of the
registration level at different depths. This is usually equal to the critical flaw size for
the application. The EPOCH 6LT allows you to adjust this registration level during a
live inspection.
NOTE
Содержание EPOCH 6LT
Страница 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Страница 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Страница 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Страница 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Страница 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Страница 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Страница 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Страница 109: ...DMTA 10084 01EN Rev 2 November 2018 Calibration 99 Figure 8 7 Calibration Successfully completed...
Страница 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Страница 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Страница 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Страница 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Страница 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Страница 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...