75
Problem
Cause
Remedy
Page
Auto scale switching does not work.
The objective magnifications are not
input or set erroneously.
Input the magnifications of the objec-
tives in use.
67
Functions interlocked with microscope
operations do not work.
Cables are connected improperly.
Connect the cables of the microscope,
control box (BX3-CBH, etc.) and control
box D21-CB correctly.
The control box is not recognized.
Set the main switches of the control
box (BX3-CBH, etc.) and the control box
D21-CB, and set the switches to ON in
the order of: 1) control box (BX3-CBH,
etc.) ; 2) control box D21-CB.
64,65
67
When the power is switched OFF, the
MENU and scale settings return to the
default settings.
The control box D21-CB was shut down
forcefully (the main switch was kept
pressed) when the power was OFF.
Do not keep the main switch pressed
when the power is OFF. A short press
of the main switch turns off the power.
18