Chapter 6 Lamp Replacement
41
EVIS EXERA XENON LIGHT SOURCE CLV-160
• Never install a lamp that has not been approved by Olympus.
The use of a non-approved lamp can cause damage to the
light source and ancillary equipment, malfunction or fire.
• Do not apply shock, excessive force or scratches to the lamp.
It could break the glass and/or shorten the lamp life due to
the high internal pressure of the lamp.
• When disposing of the used lamp, make sure to cut the
projection part on the “+” side of the examination lamp by
using a nipper and discharge the internal gas (see Figure
6.1). Otherwise, the glass could break due to the high
internal pressure of the lamp.
• When cutting the projection part on the “+” side of the
examination lamp, wear appropriate protection equipment all
the time and be sure to cover the lamp with cloth. Otherwise,
the projection part could scatter in all direction and cause
injury to people around. The protective equipment may
include eye wear, face mask and gloves.
Figure 6.1
• Do not touch the glass surface of the lamp, filter or reflector.
Natural skin moisture from your fingers can cause cracks and
damage the light source.
• Handle the lamp carefully. Otherwise, the lamp may be
damaged, resulting in equipment failure.
• When replacing the examination lamp, use a clean lint-free
cloth to wipe off residual heat compound from the heat sink. If
the heat compound is not wiped off completely, the lamp’s
heat efficiency will be impaired and the examination lamp life
will be shortened significantly.
Projection part
Pin
Examination lamp