Chapter 4 Inspection
EVIS EXERA XENON LIGHT SOURCE CLV-160
4.4
Inspection of air and water feeding
When using a water container to feed water from the endoscope, inspect the air
and water feeding functions as follows.
1.
Confirm that one of the airflow indicators (“L”, “M”, and “H”) is lit. If no
indicator is lit, press the airflow regulator ON switch once. Then confirm that
a beep is heard and one of indicators (“L”, “M”, and “H”) is lit (see Figure
4.7).
• When the light source is turned OFF, the airflow regulator
setting is automatically stored and recalled when the light
source is ON again.
• The airflow regulator has been set to OFF before shipment.
Press the airflow regulator ON switch once to set the airflow
regulator to “L”.
Figure 4.7
2.
With the airflow regulator set to “L”, “M” or “H”, immerse the distal end of the
connected endoscope in sterile water and operate the endoscope’s air
feeding function as described in the endoscope’s instruction manual.
Airflow indicators
Airflow regulator
ON switch
Airflow regulator
OFF switch