16
Chapter 3 Installation and Connection
EVIS EXERA XENON LIGHT SOURCE CLV-160
3.3
Connection of an endoscope
Before connecting the endoscope’s light guide connector to
the light source, make sure that it is perfectly dry. If the
endoscope’s light guide connector is wet (after disinfecting,
for example), wipe and dry the exterior as described in the
endoscope’s reprocessing manual. Otherwise, electrical
shock or equipment damage can result.
Insert the light guide connector into the output socket on the front panel of the
light source until it clicks into place (see Figure 3.5).
Figure 3.5
3.4
Connection of the OES video system
• Set the auto/manual brightness selector to “AUTO” to
prevent burns caused by close observation. So the
brightness is adjusted automatically. If using manual
brightness adjustment, set the brightness to the middle or
lower level and never observe close to mucous membranes
for a long time.
• When using rigidscope, set the auto/manual brightness
selector to “AUTO” to prevent burns caused by close
observation. So the brightness is adjusted automatically. If
using manual brightness adjustment, set the brightness to
the middle or lower level and never observe close to mucous
membranes for a long time.