36
The diagram below shows the sequence of assembly of the modules. The numbers indicate the order of assembly.
The module numbers shown in the following diagram are merely the typical examples. For the modules with which the
module numbers are not given, please consult your Olympus representative or the latest catalogues.
# When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts
or touching glass surfaces.
Assembly steps enclosed in will be detailed on the subsequent pages.
}All assembly operations are possible by using the Allen screwdriver ( ) provided with the microscope frame.
However, the reflected light illuminator or transmitted light arm should be attached using the Allen wrench provided with the
illuminator or arm to tighten the clamping screws. (To assure the performance, please have your dealer assemble the
illuminator or arm.)
The filter turret and cross stage are to be cleaned respectively using the special tools provided with them.
(Note) For the detailed assembly procedures of the reflected light fluorescence system and the TH4 power supply unit, refer to
their instruction manuals.
ASSEMBLY
9-1 Assembly Diagram
Eyepieces
WHN10X (FN 22)
35WHN10X (FN 22)
Trinocular tube
U-TR30-2 (FN 22)
U-ETR3 (FN 22)
Revolving arm
Revolving
nosepiece
Objectives
(See next page.)
Intermediate
attachment
U-ECA
U-FWO
U-KPA, etc.
Transmitted light arm
BX-ARM
Reflected light fluo-
rescence illuminator
BX-URA2
BX-RFA
Reflected light mercury bulb
lamp housing
U-LH100HG
U-LH100HGAPO
Cross stage
IX-SVL2
Stage center plate
Microscope frame
BX51WIF
Power supply unit
U-RFL-T
Fixed stage
adapter
WI-FSH
Halogen lamp housing
U-LH100-3
U-LH100IR
Condenser clamping knob
Lamp socket clamping screw
Filter turret
Filter frame
reinforcing ring
Extension cord
U-RMT
Power supply unit
TH4
Hand switch
TH4-HS
Condenser
WI-UCD
WI-DICD
WI-OBCD
U-UCD8
U-SC3