DMTA-10004-01EN, Rev. D, November 2016
Using Special Functions
123
Figure 8
‑
7 Setting up the STANDARD alarm
4.
When
ALARM MODE
is set to
STANDARD
or
B
‑
SCAN
, set the
LOW ALARM
and the
HIGH ALARM
values.
OR
When
ALARM MODE
is set to
PREVIOUS THK
:
a)
With
PREVIOUS THK MODE
set to
ABSOLUTE DIFF
[absolute
differential], set the
ABSOLUTE LOSS
and the
ABSOLUTE GROWTH
values.
b)
With
PREVIOUS THK MODE
set to
% DIFF
[percent differential], set the
% LOSS
and the
% GROWTH
values.
OR
When
ALARM MODE
is set to
REDUCTION RT
, set the
YELLOW ALARM
and
the
RED ALARM
values.
5.
Press
[MEAS]
to return to the measurement screen.
Alarm reference values that were entered in one unit system are displayed as the
equivalent value when the alternate units are selected.
NOTE
Содержание 38DL PLUS
Страница 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Страница 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Страница 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Страница 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Страница 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Страница 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Страница 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Страница 221: ...DMTA 10004 01EN Rev D November 2016 Using the Datalogger 209 f Press MEAS to exit the minimum review ...
Страница 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Страница 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Страница 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Страница 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Страница 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Страница 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Страница 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...