DMTA-10004-01EN, Rev. D, November 2016
Using Advanced Gaging Features
137
— Reduce the gain value when noise peaks are above 20 % of the waveform
display height.
This prevents the gage from detecting a noise peak rather than the correct
back-wall echo.
3.
Press
[GAIN]
again to restore the last previously set gain.
4.
Press
[GAIN]
again to return to the automatically adjusted echo height mode.
The default automatic gain mode is indicated by a blank
Gain
field.
10.2 Adjusting the Extended Blank with Dual Element Transducers
Olympus recommends using the extended blank parameter only when you are an
experienced operator who thoroughly understands the acoustic properties of the
material being measured. Incorrect use of the extended blank can cause the gage to
misread areas of thin material.
Normally, the 38DL PLUS searches for echoes down to nearly zero thickness.
However, some special circumstances such as a high degree of near surface corrosion,
aluminum material, enclosed flaws, or laminations can generate echoes that the
instrument may falsely detect as the low thickness. When these echoes are larger than
the sought after back-wall echo, the manual gain adjustment (see “Adjusting the Gain
with Dual Element and E110 EMAT Transducers” on page 135) cannot prevent this
false detection. However, the extended blank parameter allows you to define an early
period in which the 38DL PLUS will not perform echo detection, and therefore,
prevent erroneous measurements.
To use the extended blank
1.
Press
[WAVE ADJ]
.
The waveform adjustment parameter and its value appear on the measurement
screen (see Figure 10-2 on page 138).
2.
If needed, use the
[
]
and
[
]
keys to select
EXT BLANK
.
The extended blank becomes active, but initially, its value is zero. The gage
remains in the measurement screen.
CAUTION
Содержание 38DL PLUS
Страница 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Страница 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Страница 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Страница 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Страница 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Страница 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Страница 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Страница 221: ...DMTA 10004 01EN Rev D November 2016 Using the Datalogger 209 f Press MEAS to exit the minimum review ...
Страница 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Страница 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Страница 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Страница 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Страница 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Страница 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Страница 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...