DMTA-10004-01EN, Rev. D, November 2016
Configuring the Instrument
129
The
AGC
function works well for most standard thickness gage applications and is by
default turned on. In some thickness applications, the receiver gain is set at or near its
maximum value. In such cases, turn off the AGC function to prevent reading
instability.
12. For single element transducers only, set
AVERAGING
to
OFF
to turn thickness
averaging off, set
AVERAGING
to
ON
to perform a running average of the last
5 thickness readings, or set
AVERAGING
to
On
‑
QBar
to display a quality
measurement Q-bar below the measurement screen indicating the stability of the
averaged reading.
13. For dual element transducers only, set
GAIN OPT
to
ON
to activate the gain
optimization during the zero and material velocity calibrations (see “Material
Sound Velocity and the Zero Calibrations” on page 71 for details).
14. For dual element transducers only, set
VPATH CAL ENABLE
to
ON
to activate
the V-Path calibration option when you press
[CAL VEL]
Correction Curve for a Nonstandard Dual Element Transducer” on page 216 for
details).
15. Press
[MEAS]
to return to the measurement screen.
9.2
Configuring System Parameters
The
SYSTEM
screen, allows you to configure many 38DL PLUS system parameters.
To configure system parameters
1.
From the measurement screen, press
[SETUP MENU]
.
2.
In the menu, select
SYSTEM
.
The
SYSTEM
screen appears (see Figure 9-2 on page 130).
TIP
Содержание 38DL PLUS
Страница 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Страница 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Страница 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Страница 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Страница 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Страница 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Страница 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Страница 221: ...DMTA 10004 01EN Rev D November 2016 Using the Datalogger 209 f Press MEAS to exit the minimum review ...
Страница 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Страница 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Страница 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Страница 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Страница 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Страница 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Страница 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...