54
OLC(T)100
Technical manual
Type of gas
Measurement
range
Accuracy
T
50
/ T
90
(s)
Freon R23
Freon R141 b
Freon R143 a
Freon R404 a
Freon R507
Freon R410 a
Freon R32
Freon R227
Freon R407 c
Freon 408 a
1 % vol
2000 ppm
2000 ppm
2000 ppm
2000 ppm
1000 ppm
1000 ppm
1 % vol
1000 ppm
4000 ppm
70% FS)
25 / 50
Ethanol
Toluene
Isopropanol
2-butanone ( MEK)
Xylene
500 ppm
500 ppm
500 ppm
500 ppm
500 ppm
+/- 15% (from 20 to
70% FS)
25 / 50
Infrared head (OLCT100 XP-IR)
❧
Measurement range:
0–100% LEL (explosive gases)
0–5% CO
2
(carbon dioxide)
❧
Measurement principle:
Infra-red absorption
❧
Accuracy:
- CO
2
version:
+/- 3% of full-scale at mid-scale (20°C)
- LEL
version: +/- 5% of full-scale at mid-scale (20°C)
❧
Temperature range:
-25 to +55 °C
❧
Relative humidity:
0 to 95 % RH (non-condensing relative humidity)
❧
Pressure:
Partial pressure measurement (the measurement
changes with pressure)
❧
Response time:
- CO
2
version: T
50
11 s and T
90
30 s
- LEL
version: T
50
11 s and T
90
30 s
❧
Lifetime (typical):
48 months
❧
Storage conditions:
4–20°C
10–60% RH
1 bar ± 10%
6 months maximum
❧
Warm-up time (max):
2 hours to first switching on power
Содержание OLC 100
Страница 1: ...P N NPO100GB rev C...
Страница 45: ...9 Declaration of EC conformity 45...
Страница 46: ...46 OLC T 100 Technical manual Chapter 10 Technical specifications...
Страница 50: ...50 OLC T 100 Technical manual Mark on sensor VQ1 4F poison resistant sensor Figure 25 mark on VQ1 sensor...
Страница 62: ...62 OLC T 100 Technical manual...
Страница 63: ...Annexe Ordering information 63...