Annexe – Ordering information
49
Weight :
❧
OLC100: 0.950 kg.
❧
OLCT100 XP HT: 1.8 kg.
❧
OLCT100 XP LEL: 1.0 kg.
❧
OLCT100 XP IR: 1.1 kg.
❧
OLCT100 XP EC: 1.1 kg.
❧
OLCT100 XP SC: 1.1 kg.
Materials:
Epoxy painted aluminum Optional 316 stainless
Catalytic head (OLCT100 XP)
Common characteristics
❧
Measurement range
0–100% LEL
❧
Measurement principle:
catalytic filaments
❧
Accuracy:
see table below
❧
Temperature range:
see table below
❧
Relative humidity:
0 to 95% RH (non-condensing relative humidity)
❧
Pressure:
atmospheric ± 10%
❧
Response time:
T
50
= 6 seconds. T
90
= 15 seconds for Methane
❧
Lifetime (typical)
48 months
❧
Storage conditions:
-50 to 70°C, 20 to 60% RH, 1 bar ± 10%, 6 months
maximum
❧
Warm-up time (max)
2 hours to first switching on power
Specific characteristics
Type of sensors
Accuracy
Operating
temperature
range
Anti-poison sensor
4F (unmarked
sensor)
1
% LIE between 0- 70 %LIE
2% of the measurement between 71 and
100% LIE
-20 to +70°C
VQ1 sensor
(sensor with
identifying mark)
1
% LIE between 0- 70 %LIE
OLCT100 : 2% of the measurement between 71
and 100% LIE
OLC 100 : 5% of the measurement between 71
and 100 % LIE
-40 to +70°C
VQ1 sensor, high
temperature
assembly
1
% LIE between 0-70%LIE
2% of the measurement between 71 and
100% LIE
-20 to +200°C
Содержание OLC 100
Страница 1: ...P N NPO100GB rev C...
Страница 45: ...9 Declaration of EC conformity 45...
Страница 46: ...46 OLC T 100 Technical manual Chapter 10 Technical specifications...
Страница 50: ...50 OLC T 100 Technical manual Mark on sensor VQ1 4F poison resistant sensor Figure 25 mark on VQ1 sensor...
Страница 62: ...62 OLC T 100 Technical manual...
Страница 63: ...Annexe Ordering information 63...