Annexe – Ordering information
53
Type of
gas
Accuracy
(ppm)
Life
(months)
Response
time
T
50
/ T
90
(s)
Storage
conditions
Warm-up
time max
(h)
NO
2
+/- 0,8
24
30/60
(1)
12
O
2
0,4 % vol
(15 to 22 % O
2
)
28
6-15
(1)
Aucun (3)
PH
3
+/- 0,05
18
30/120
(1)
1
SiH
4
+/- 1
18
25/120
(1)
1
SO
2
+/- 0,7
(range 0-10)
36
15/45
(1)
1
(1)
4 – 20 °C
20 – 60 % RH
1 bar ± 10 %
6 months maximum
(2)
4 – 20 °C
20 – 60 % RH
1 bar ± 10 %
3 months maximum
(3)
If cartridge is mounted in the transmitter
Semiconductor heads
(OLCT100 XP)
Common characteristics
❧
Measurement principle:
semiconductor
❧
Temperature range:
-20°C to +60°C
❧
Relative humidity:
20 to 95% RH (non-condensing relative humidity)
❧
Pressure:
atmospheric ± 10%
❧
Lifetime (typical):
36 months
❧
Storage conditions:
-20 to 50 , 20 to 60% RH, 1 bar ± 10%, 6 months
maximum
❧
Warm-up time (max):
4 hours to first switching on power
Type of gas
Measurement
range
Accuracy
T
50
/ T
90
(s)
Methyl chloride
CH
3
Cl
Methylene chloride
CH
2
Cl
2
500 ppm
500 ppm
+/- 15% (from 20 to
70% FS)
25/50
Freon R12
Freon R22
Freon R123
FX56
1 %vol
2000 ppm
2000 ppm
2000 ppm
+/- 15% (from 20 to
70% FS)
25 / 50
Freon R134 a
Freon R142 b
Freon R11
2000 ppm
2000 ppm
1 % vol
+/- 15% (from 20 to
70% FS)
+/- 15% (from 20 to
25 / 50
Содержание OLC 100
Страница 1: ...P N NPO100GB rev C...
Страница 45: ...9 Declaration of EC conformity 45...
Страница 46: ...46 OLC T 100 Technical manual Chapter 10 Technical specifications...
Страница 50: ...50 OLC T 100 Technical manual Mark on sensor VQ1 4F poison resistant sensor Figure 25 mark on VQ1 sensor...
Страница 62: ...62 OLC T 100 Technical manual...
Страница 63: ...Annexe Ordering information 63...