7
–
Menus
39
and sensitivity or
Esc
to exit the procedure.
2. The detector is calibrated.
5. Sensor exchange
This function reboots the parameters (rate of wear, calibration date, internal
parameters corresponding to the 4-20mA range, etc.) from the selected
detector(s) following or in view of a change of cell.
Detector Selection
1. Select the detector(s) to be rebooted with the help of the
Previous
detector
, and
Next detector
keys and press
Select.
Initialization of the Detector
1. Press Escape to reset the selected cells
2. Proceed with the cell replacement and then calibrate locally the
corresponding detectors
3. Proceed with the calibration from the
MX 32
via the menus “1 Sel
detectors”, “2 recording”, "End recording” and “4 validation” in order to
save data calibration (wear rate, date of calibration, response time, etc.).