6
–
COM2100
software
57
Alarm relay configuration and logic inputs
■
Set relay output and the logic input by
checking the different boxes (Figure 45).
Reference 1
Alarm relay output settings. In this example
all CO, O2 and LEL alarms will trigger the
relay output.
Reference 3
To set the BM 25A/AW modes when logic
input is activated (set to 1).
-
Alarm relay
: internal relay output is
activated
-
Alarm 1
: BM 25A/AW will turn in Alarm 1
mode
-
Alarm 2
: BM 25A/AW will turn in Alarm 2
mode
-
Remote alarm
: BM 25A/AW will turn in
Alarm Transfer mode
Reference 4
Alarm acknowledgment
-
Local acknowledgment
: alarms must be
acknowledged from the gas monitor
keyboard.
-
Remote acknowledgment
: when on a
network (wired or wireless), local alarms
can be acknowledged from another BM
25A/AW.
■
Click
OK
to confirm.
Figure 45: alarm output relay and
logic input settings
Содержание BM 25A/AW
Страница 1: ...BM 25A AW Area Gas Monitor Part Number NP25AEN Version C 4 User Manual ...
Страница 10: ...10 BM 25A AW Instruction Manual Product Overview Figure 2 Product Overview ...
Страница 18: ...18 BM 25A AW Instruction Manual ...
Страница 34: ...34 BM 25A AW Instruction Manual ...
Страница 70: ...70 BM 25A AW Instruction Manual ...
Страница 75: ...9 Special Instructions 75 CSA Control Drawing for BM 25A AW Connections ...
Страница 78: ...78 BM 25A AW Instruction Manual ...
Страница 79: ...10 Declaration of EC Conformity 79 ...