4
– Wireless Version
39
Main BM 25AW
Secondary BM 25AW
Controller
Case
Cause
Flash
Siren
Flash
Siren
Channel
status
1.
Fault (no
communication
at all, sensor
fault, low
battery, etc.)
Steady
Yes
No
No
Fault
2.
Alarm 1
1 Hz
1 Hz
0.5 Hz
0.5 Hz
Alarm 1
3.
Alarm 2
2 Hz
2 Hz
0.5 Hz
0.5 Hz
Alarm 2
Tableau 2 : Controller Mode
– Table of events
Start-up
From the Maintenance menu (see Chapter 5), choose Wireless. Leave it to 'OFF' if you
do not want to activate the radio function. Select '
BM 25
' or 'Controller' according to the
chosen operation mode (see above).
Figure 25 : Wireless mode screenshots
■
In BM 25 mode
: only the network ID (Channel) must be set between 0 and 15.
Address number (Slv Number) is not editable and is set to 'XXX'. In this mode, it is
not necessary to assign an address as the network is automatically built in by using
the MAC
(*)
addresses of each device.
■
In Controller mode
: set the BM 25AW address
Slv Nb
between 1 and 30 max. and
the network ID
Channel
between 0 and 15.
(*) MAC (Media Access Control): unique identifier assigned to network interfaces. Each BM 25AW
has its own unique MAC address.
WARNING
Содержание BM 25A/AW
Страница 1: ...BM 25A AW Area Gas Monitor Part Number NP25AEN Version C 4 User Manual ...
Страница 10: ...10 BM 25A AW Instruction Manual Product Overview Figure 2 Product Overview ...
Страница 18: ...18 BM 25A AW Instruction Manual ...
Страница 34: ...34 BM 25A AW Instruction Manual ...
Страница 70: ...70 BM 25A AW Instruction Manual ...
Страница 75: ...9 Special Instructions 75 CSA Control Drawing for BM 25A AW Connections ...
Страница 78: ...78 BM 25A AW Instruction Manual ...
Страница 79: ...10 Declaration of EC Conformity 79 ...