3
– Operation
23
Chapter 3
| Operation
Start-up
When starting up the BM 25A/AW for the first time or after a period of
inactivity longer than one month, it is recommended to proceed a charge-
discharge cycle. Moreover, keep in mind that all portable gas monitors
must be tested with gas before each day of use.
Switching on the instrument
■
Press the
Enter
button (ref. 1).
■
Before displaying current measurements, the
BM25 A/AW performs visual and audible tests
during a few seconds and then displays:
- The OLDHAM SIMTRONICS logo,
- The software revision and the serial number,
- The alarm thresholds set for each measurement
channel.
■
on page 24.
Note: During normal operation, the BM 25A/AW
flashes every 2 minutes to indicate that it operates
correctly. This confidence flash can be canceled and
the frequency can be changed using the
COM 2100
software; see page 51.
Figure 13: start-up in
standard mode by pressing
the Enter button
Selection of the flammable gas on start-up
On start-up, it is possible to change the calibration gas. This option allows the user to
measure a different flammable gas (acetone, methanol, etc.) from that which was
originally used (methane or pentane) to calibrate the instrument. This action allows the
instrument software to compensate and display more accurate readings.
Note:
Accuracy for the re-selected gas type is +/- 15%.
Содержание BM 25A/AW
Страница 1: ...BM 25A AW Area Gas Monitor Part Number NP25AEN Version C 4 User Manual ...
Страница 10: ...10 BM 25A AW Instruction Manual Product Overview Figure 2 Product Overview ...
Страница 18: ...18 BM 25A AW Instruction Manual ...
Страница 34: ...34 BM 25A AW Instruction Manual ...
Страница 70: ...70 BM 25A AW Instruction Manual ...
Страница 75: ...9 Special Instructions 75 CSA Control Drawing for BM 25A AW Connections ...
Страница 78: ...78 BM 25A AW Instruction Manual ...
Страница 79: ...10 Declaration of EC Conformity 79 ...