APPENDIX C SOFTWARE SERVICE TOOL INSTRUCTIONS
C.2 Configuring the Scale (after replacing the PCB or Load Cell)
1. When communications are reliable, return to the Config tab. Press the
Read
button (E)
and the data will be read from the scale.
2. If the PC Board has been replaced, the serial number will not match the balance. To
configure the scale, enter its serial number in the blank (C), then click
Scale
in the
Serial
Number
box (A), and click
OK
(D). When the
OK
button highlights again, click
Update
(F). Turn the scale off and then back on to save the data.
A
B
C
D
G
Figure C-4. The Config Tool’s Configuration Screen.
E
F
3. If the Load Cell was replaced, follow the same procedure as above, but enter the Load
Cell serial number in the blank (C), and click the
Load Cell
(B) radio button.
4. The bar (G) is a graphical measure of the Load Cell output currently being used.
5. Turn the scale off, perform a Service Calibration (see Appendix B), followed by
Operational and Performance tests (Chapter 4).
Traveler™ Series Service Manual
C-3
Ohaus Corporation www.ohaus.com
Содержание TA152
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 23: ...CHAPTER 3 MAINTENANCE PROCEDURES Ohaus Corporation www ohaus com 3 4 Traveler Series Service Manual...
Страница 37: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 8 Traveler Series Service Manual...
Страница 39: ...APPENDIX A STANDARD CALIBRATION Ohaus Corporation www ohaus com A 2 Traveler Series Service Manual...
Страница 43: ...APPENDIX B THE SERVICE MENU Ohaus Corporation www ohaus com B 4 Traveler Series Service Manual...
Страница 48: ......
Страница 49: ...80251719 P N 80251719 SERVICE MANUAL TRAVELER SERIES SCALES...