3.3
Set calibration standard
After the MTC temperature value setting, the meter goes to setup the calibration standard.
Use
or
buttons to select the calibration standard.
Press button-
to confirm your setting.
Continue with calibration standard setting by following instructions or press button-
to go
back to the measurement screen.
3.4
Set temperature correction coefficient
After the calibration standard setup, the meter enters the temperature correction coefficient
setting. Use
or
buttons to increase or decrease the temperature correction value (0.00
to 10.00 %/°C). Press button-
to confirm your setting. Continue with reference temperature
setting or press button-
to go back to the measurement screen.
Note:
If you set the temperature correction coefficient value as 0 %/°C, it means there will be
NO
Temperature Compensation
for the conductivity measurement. The meter displays the real
conductivity value at current temperature.
3.5
Set TDS factor
After selecting the reference temperature, the current TDS factor blinks. Use
or
buttons to increase or decrease the value. Press
to confirm your setting. The meter will exit
automatically to the measurement screen.