EN 14                                                                                             Ranger

 4000 Count Series    

 
 
 

 

Reset

 [

reset

]

  

Reset the Setup menu to factory defaults. 
 

NO 

= not reset 

 

YES  

= reset 

 

Dual Scale [

d.scale

]

   

Set the status of the second scale input (Scale 2). 
 

OFF 

= disabled 

 

ON 

= enabled 

 

Cap2 [

CAP2

]

  

Set the capacity of Scale 2. 
 

1…9999 

 

Grad2 [

GrAd2

]

  

Set the readability of Scale 2. 
 

0.00005…0.5 

 

Power on unit

 [

pwWr.UN

]

  

Set the unit of measure displayed at startup. 
 

AUTO 

= last unit in use when turned off 

 

kg 

= kilograms 

 

= grams 

 

lb 

= pounds 

 

oz 

= ounces 

 

lb:oz 

= pound ounces 

 

Zero Range

 [

ZErO

]

  

Set the percentage of scale capacity that may be zeroed. 
 

2% 

= zero up to 2 percent of capacity 

 

10% 

= zero up to 10 percent of capacity 

 

Auto Opt [

A.OPt

]

  

Set the Automatic Optimization. 
 

OFF 

= disabled 

 

ON 

= APW automatically optimized  

 

Auto Tare

 [

A.tArE

]

  

Set the automatic tare functionality.  
 

OFF 

= Automatic Tare is disabled 

 

ON 

= the first stable gross weight is tared 

 

ON-ACC  = stable gross loads within the accept limits are tared (in Check weighing mode) 

 

Beeper Opt [

bp.OPt

]

  

Set whether or not the beeper sounds when APW has been optimized. 
 

OFF 

= disabled 

Содержание RC41M15-M

Страница 1: ...Ranger 4000 Count Series Instruction Manual Manual_RC41_V1...

Страница 2: ......

Страница 3: ...azard Alternating Current 1 2 Safety Precautions Please follow these safety precautions Verify that the AC input voltage printed on the data label matches the local AC power supply Only connect the po...

Страница 4: ...al pan 2 3 Selecting the Location Use the scale on a firm steady surface Avoid locations with excessive air current vibrations heat sources or rapid temperature changes Allow sufficient space around t...

Страница 5: ...eration the battery symbol indicates the battery charge level The scale will automatically turn off when the batteries are fully discharged For maximum operating time the battery should be charged at...

Страница 6: ...s2 Off Turn scale off Units Change the weighing unit Switch between Check Weigh Check Count and Off Menu Enter user menu View the Audit Trail event counters extended press Menu Function Short Press Ye...

Страница 7: ...ss and hold for more than 2 seconds TABLE 3 2 LCD Symbols Item Description Item Description 1 NET symbol 10 Low Sample Weight Symbol 2 Center of Zero symbol 11 Auto Optimization Symbol 3 Negative symb...

Страница 8: ...active weighing mode To turn the scale off press and hold the On Zero Off button until OFF is displayed 3 3 Counting Ranger Count 4000 has three display areas Weight Piece Weight and Count informatio...

Страница 9: ...teger 5 Press the Sample key will be displayed in both the piece window and the count window 6 The established APW and count will be displayed Entering a known APW 1 Press the APW key will be displaye...

Страница 10: ...on the pan or in the container If the sample weight is under the target weight range the yellow LED will light If the sample is within the target weight range the green LED will light If the sample is...

Страница 11: ...ly a different ID number may be selected using the keypad and then pressing the ID key Note Maximum 30 records from 0 to 29 Retrieving a record 1 Enter the ID number using the keypad and press the ID...

Страница 12: ...nter the ID number using the keypad and press the ID key reCall and lib are displayed in the weight and piece weight windows respectively 2 Press the C key delete is displayed in the weight window Pre...

Страница 13: ...o On Some units may not be available in all models When LEGAL FOR TRADE is set to ON LFT Switch in locked position the menu settings are affected as follows Calibration C A L menu is not accessible Du...

Страница 14: ...te weighing results Before performing the calibration be sure to have the appropriate calibration weights as listed in table 4 1 Ensure that the LFT switch calibration lock is set to unlocked position...

Страница 15: ...m 0 to 31 are available with 12 being the default Refer to table 4 2 to determine the GEO factor that corresponds to your location End Cal End Advance to the next menu or return to the top of the curr...

Страница 16: ...f kg kilograms g grams lb pounds oz ounces lb oz pound ounces Zero Range ZErO Set the percentage of scale capacity that may be zeroed 2 zero up to 2 percent of capacity 10 zero up to 10 percent of cap...

Страница 17: ...utton is pressed OFF no sound ON sound Accumulation aCCUmM Set the accumulation functionality OFF accumulation disabled AUTO automatic accumulation MAN manual accumulation End Setup End Advance to the...

Страница 18: ...anges Auto off A OFF Set the automatic shut off functionality OFF disabled 1 powers off after 1 minute of no activity 5 powers off after 5 minute of no activity 10 powers off after 10 minute of no act...

Страница 19: ...time the stability criteria are met INTER printing occurs at the defined interval ACCEPT printing occurs continuously CONT printing occurs each time the display is within the Checkweigh accept range a...

Страница 20: ...nted Info Set the status OFF disabled ON the reference information is printed APW Set the status OFF disabled ON enabled Accu Set the status OFF disabled RESULT the Accumulation result is printed ALL...

Страница 21: ...l Notes For Windows 7 setup contact Ohaus If the weighing value is a negative number set the target cell in TEXT format Otherwise Excel will not distinguish it as a negative number Please do not use t...

Страница 22: ...ng ON OFF XON XOFF software handshaking Alternate command Alt CmM Define command character for the Print Tare and Zero commands Print Set the alternate command character for Print A to Z Tare Set the...

Страница 23: ...et the status OFF COM menu is not locked ON COM menu is locked End Lock End Advance to the next menu or return to the top of the current menu 4 9 End Menu Press Yes to advance to the Calibration menu...

Страница 24: ...EN 22 Ranger 4000 Count Series Figure 4 1 Setup for the Weigh Below Hook Weigh Below Hook Access...

Страница 25: ...2 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15 15 14 14 13 13 12 44 32 45 38 18 17 17 16 16 15 15 14 14 13 13 45 38 46 45 18 18 17 17 16 16 15 15 14 14 13 46 45 47 51 19...

Страница 26: ...s explained in Section 3 9 3 Set the switch to Locked See figure 5 1 The Menu Lock switch limits changes to the Cal Setup Readout Unit and Print menus The switch in type approved models may set some s...

Страница 27: ...Ranger 4000 Count Series EN 25 Figure 5 1 Sealing Unlocked Locked with Paper Seal Locked with Wire Seal Paper Seal Wire Seal Lock Switch...

Страница 28: ...wer On Error Weight reading below Power On Zero limit Err 8 3 Over Range Error Weight reading exceeds Overload limit Err 8 4 Under Range Error Weight reading below Underload limit Err 8 5 Tare out of...

Страница 29: ...0 lb x 0 001 lb 480 ox x 0 02 oz 30 kg x 0 001 kg 30000 g x 1 g 60 lb x 0 002 lb 960 oz x 0 04 oz Maximum Displayed Resolution 1 30000 Minimum Recommended Sample Weight 2g 0 004lb 4g 0 01lb 10g 0 02lb...

Страница 30: ...r unstable CP Continuous Print SP Print on Stability xS 0S Turn off Stable Only menu item and allow unstable print 1S Turn on Stable Only menu item and only print stable print xP Interval Print x Prin...

Страница 31: ...he pin connections are as follows Active pins PIN 2 TXD PIN 3 RXD PIN 5 Ground Figure 7 2 RS232 Pins 7 4 Printouts The following sample print outs are generated by the Print button P Command or altern...

Страница 32: ...iate left of most significant character if negative Unit Up to 5 characters left justified If the Unit in the Print Content menu was set to OFF the unit will be removed in the weight string and replac...

Страница 33: ...turers work It comprises all tests according to the adopted European standard EN 45501 1992 paragraph 8 2 2 If national regulations limit the validity period of the verification the user of the weighi...

Страница 34: ...e has been designed for safe removal of the batteries at end of life by a waste treatment facility Please dispose of this product in accordance with local regulations at the collecting point specified...

Страница 35: ...has been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other th...

Страница 36: ...edor del mundo Avec des bureaux dans le monde entier Weltweite Gesh ftsstellen Con uffici in tutto il mondo www ohaus com 30037454 P N 30037454A 2015 OHAUS Corporation all rights reserved todos los de...

Отзывы: