Ranger
™
4000 Count Series EN 13
Zero2
[
ZErO
]
Initiates a 2
nd
scale zero calibration.
Span1
[
SpaN
]
Initiates a span calibration procedure (zero and span). A span calibration is important when
initially setting up the scale.
Span2
[
SpaN2
]
Initiates a 2
nd
scale span calibration procedure.
Lin1 [
LIN
]
Initiates a linearity calibration procedure (zero, mid-point and span).
Lin2 [
LIN2
]
Initiates a 2
nd
scale linearity calibration procedure.
GEO [
GEO
]
Geographical Adjustment Factor (GEO) is used to adjust the calibration based on the current
location. Settings from 0 to 31 are available with 12 being the default. Refer to table 4-2 to
determine the GEO factor that corresponds to your location.
End Cal [
End
]
Advance to the next menu or return to the top of the current menu.
4.3 Setup Menu
Enter this menu to set scale parameters.
Reset:
no
, yes
Dual Scale:
off
, on
Cap2:
1
…9999
Grad2:
0.00005…0.5
Power on unit:
auto
, kg, g, lb, oz, lb:oz
Zero Range:
2%,
10%
Auto Opt:
off,
on
Auto Tare:
on,
off
, on-acc
Beeper Opt:
off,
on
Beeper Signal:
off
, accept, under, over, under-over
Beeper Key:
off
, on
Accumulation:
off
, auto, manual
End Setup:
Exit menu
Zero2:
Perform
Span1:
Perform
Span2:
Perform
Linearity1:
Perform
Linearity2:
Perform
Geographic
Adjustment:
Set 0
…
12
…31*
End Calibration: Exit menu
*Bold