CHAPTER 4 TESTING
Navigator Series Service Manual
19
Ohaus Corporation www.ohaus.com
4 TESTING
Before and after servicing a Navigator scale, an operational test and various performance tests should
be made to confirm that the scale meets specifications. Turn the scale on and allow it to warm up for
at least one hour before performing these tests.
NOTE:
Make sure the test area is free from drafts and that the scale rests on a level and vibration-free
surface.
4.1 TEST MASSES REQUIRED
The masses required to test the Ohaus Navigator scales must meet the requirements of ASTM
Class 4 or OIML F2 Tolerance. The mass values are listed in Table 4-1.
TABLE 4-1
CALIBRATION MASS VALUES
Model
Span
Calibration mass
Linearity
Calibration masses
NV222
200g
100g, 200g
NV422
200g
200g, 400g
NV622
300g
300g, 600g
NV221
200g
100g, 200g
NV621
600g
300g, 600g
NV1201
500g
500g, 1kg
NV2201
1kg
1kg, 2kg
NVT1601M
1kg
1kg,1.5kg
NVT2201
1kg
1kg, 2kg
NVT4201
2kg
2kg, 4kg
NVT6201
5kg
3kg, 6kg
NVT2200
1kg
1kg, 2kg
NVT3200M
2kg
2kg, 3kg
NVT6200
5kg
3kg, 6kg
NVT6400M
5kg
3kg, 6kg
NVT12000
5kg
5kg, 10kg
NVT16000M
10kg
10kg, 15kg
NVT22000
10kg
10kg, 20kg
NV123
100g
50g,100g
NV223
200g
100g,200g
NV323
300g
150g,300g
NV1202
1kg
500g,1kg
NV2202
2kg
1kg,2kg
NV3202
3kg
1.5kg,3kg
NVT10201
5kg
5kg,10kg