3 6
Operating your Moisture Analyzer - Section 6
Ramp profile
For this temperature program, the temperature and time elapsed between
the start of drying and the attainment of the final temperature is set.
Using the arrow buttons, scroll to PROFILE and press the
Enter
button.
DRYING PROFILE
EXIT WITHOUT SAVING
Using the arrow buttons, scroll to RAMP and press the
Enter
button.
RAMP PROFILE
EXIT WITHOUT SAVING
Using the arrow buttons, scroll to END TEMP. Enter the desired tempera-
ture using the arrow buttons and press the
Enter
button.
Using the arrow buttons, scroll to RAMP TIME. Enter the desired ramp or
(elapsed time) using the arrow buttons and press the
Enter
button.
RAMP PROFILE
EXIT WITHOUT SAVING
Step profile
For this temperature program you define the temperature and the hold
time for each step.
Using the arrow buttons, scroll to PROFILE and press the
Enter
button.
DRYING PROFILE
Using the arrow buttons, scroll to STEP and press the
Enter
button.
EXIT WITHOUT SAVING
TEST PARAMETERS
TEST ID:
-DEFAULT-
PROFILE:
STANDARD
DRY TEMP:
180C
SWITCH OFF: AUTO 60
RESULT:
MOISTURE %
CUSTOM:
OFF
TEST PARAMETERS
TEST ID:
-DEFAULT-
PROFILE:
STANDARD
DRY TEMP:
180C
SWITCH OFF: AUTO 60
RESULT:
MOISTURE %
CUSTOM:
OFF
RAMP
END TEMP:
50
C
RAMP TIME:
3 MIN
END TEMP:
50 C
RAMP TIME:
3
MIN
STEP
Temperature profile
Final
temperature
Ambient
temperature
Ramp drying
Time
Start