3 0
Analyzer setup - Section 5
RS232
Exit without saving
Using the right arrow button, advance to EXIT WITHOUT SAVING and
press the
Enter
button. The display returns to ANALYZER SETUP.
Hardware
On the rear of the Moisture Analyzer is a 9-pin female subminiature D
connector provided for interfacing to other devices. Refer to the acces-
sory list at the rear of this manual for a proper cable.
Section 9.2 illustrates the pinout connections on the RS232 connector.
RS232
Setting handshake
Press the right arrow button to advance to HANDSHAKE. Using the up/
down arrow buttons, select the desired handshake of NONE, X ON-X OFF,
or RTS-CTS. Default setting is NONE. Press
Enter
button to save, display
advances to PRINT.
BAUD RATE:
9600
PARITY:
NONE
DATA BITS:
7
STOP BITS:
1
HANDSHAKE:
NONE
EXIT WITHOUT SAVING
BAUD RATE:
9600
PARITY:
NONE
DATA BITS:
7
STOP BITS:
1
HANDSHAKE: NONE
EXIT WITHOUT SAVING
RS232 Connection