EN-11
CW-11 Indicator
2.8 Initial Setup (Cont.)
2.8.4 Setup Menu
The CW-11 Indicator Setup Menu
must be entered the first time
must be entered the first time
must be entered the first time
must be entered the first time
must be entered the first time the Indicator is used to set the scale base
parameters to match the Indicator. Do not attempt to calibrate the Indicator
Do not attempt to calibrate the Indicator
Do not attempt to calibrate the Indicator
Do not attempt to calibrate the Indicator
Do not attempt to calibrate the Indicator before setting up the Setup
Menu. All other menus should be entered and set up the first time the Indicator is used.
Procedure
Procedure
Procedure
Procedure
Procedure
With the Indicator ON, press and hold the
G/N/T/
MENU
button until MENU is displayed. When you
release
G/N/T/
MENU
button, CAL is displayed when
the CAL jumper on the PC board is in place. When
the CAL jumper is removed, the Indicator will not
permit calibration. This jumper should be in place
initially.
Press
PRINT/
UNITS
button, SETuP is displayed.
Press
G/N/T/
MENU
button, LFTOFF is displayed.
legal for trade selections are:
'LFT ON' - LFT is ON
'LFTOFF' - LFT is OFF
'LFTCAn' - LFT is set for Canada
Press
PRINT/
UNITS
button and select either ON ,
OFF or Canada.
Press
G/N/T/
MENU
button, 0 2 is displayed. This
is the Zero 2%, 18% or 100% setting. 2% - zero
operation range is - 2% to + 2%. 18% - zero
operating range is -2% to +18%, 100% - zero
operation range is -2% to +100%.
NOTE
:
: :
:
: If LFT is ON, only 2% and 18% are available.
Press
PRINT/
UNITS
button, and select either 2%,
18% or 100%.
Press
G/N/T/
MENU
button, CAL Un kg is displayed.
This is the calibration unit setting. Selections are:
'lb' - calibration unit is lb
'kg' - calibration unit is kg.
Press
PRINT/
UNITS
button, and select either kg or lb.