EN-35
CW-11 Indicator
5.2Error Codes List
The following list describes the various error codes and which can appear on the display and the suggested
remedy.
L o B a t
L o B a t
L o B a t
L o B a t
L o B a t
Is indicated when batteries are weak. Approximately 20 minutes of operating time
remain.
Error 1
Error 1
Error 1
Error 1
Error 1
Indicates an overload condition.
Error 2
Error 2
Error 2
Error 2
Error 2
Indicates an underload condition.
Error 7
Error 7
Error 7
Error 7
Error 7
EEPROM data incorrect.
Error 14
Error 14
Error 14
Error 14
Error 14 Zero exceeds ZERO% and cannot be zeroed.
Err 21
Err 21
Err 21
Err 21
Err 21
Calibration data does not match current full scale, Grad and Cal Point settings. Settings
must be restored or the Indicator must be recalibrated using the current settings.
5.3 Service Information
If the Troubleshooting section does not resolve or describe your problem, you will need to contact an authorized
Ohaus Service Agent. For Service assistance in the United States, please call Aftermarket, Ohaus Corporation toll-
free at (800) 526-0659. An Ohaus Product Service Specialist will be available to help you.
5.4 Technical Data
Materials
Housing stainless-steel
Keypad/display overlay polyester
Ambient conditions
The technical data is valid under the following ambient conditions:
Ambient temperature
-10°C to 40C/ 14°F to113°F
Relative humidity
10%.......95%, noncondensing
Height above sea level
up to 4000m
Operability is assured at ambient temperatures between 5 and 40° C.