EN-40
Courier 7000
6.3.3 3835 Protocol
3835 Protocol
Request displayed weight
Command: W <CR>(57h, 0dh)
over capacity
(invalid data)
<LF> ^
^
^
^
^
^
^
^ <U> <U> <CR> <H1> <H2> <ETX>
under capacity
(-20d)
<LF> _
_
_
_
_
_
_
_ <U> <U> <CR> <H1> <H2> <ETX>
in Ib/oz/kg/g
(normal data)
<LF> <p> <W>
.
<W> <W> <W> <W> <U> <U> <CR> <H1> <H2> <ETX>
in Ib/oz/kg/g
<LF> <p> <W> <W>
.
<W> <W> <W> <U> <U> <CR> <H1> <H2> <ETX>
in Ib/oz/kg/g
<LF> <p> <W> <W> <W>
.
<W> <W> <U> <U> <CR> <H1> <H2> <ETX>
in Ib/oz/kg/g
<LF> <p> <W> <W> <W> <W>
.
<W> <U> <U> <CR> <H1> <H2> <ETX>
in Ib/oz/kg/g
<LF> <p> <sp> <W> <W> <W> <W> <W> <U> <U> <CR> <H1> <H2> <ETX>
in Ib:oz
<LF> <p> <W>
l
b
<sp> <W> <W>
.
<W> <W>
o
z <CR> <H1> <H2> <ETX>
in Ib:oz
<LF> <p> <W> <W>
l
b
<sp> <W> <W>
.
<W>
o
z <CR> <H1> <H2> <ETX>
in Ib:oz
<LF> <p> <sp> <W> <W> <W>
l
b <sp> <W> <W>
o
z <CR> <H1> <H2> <ETX>
Initial Zero Error <LF>
<H1> <H2> <CR> <ETX>
Request current status
Command: S <CR> (53h, 0dh)
Response
<LF>
<H1> <H2> <CR> <ETX>
Request scale to zero
Command: Z <CR> (5ah, 0dh)
Response
Scale is zeroed, no
response from scale
Unrecognized Command
Command: others
Response
<LF>
?
<CR>
Definition of the Symbols and the Bits
Symbols Used
<LF>
line feed (0Ah)
<CR>
carriage return (0Dh)
<ETX>
end of text (03h)
<sp>
space (20h)
<p>
polarity “-” or “ ” (2Dh or 20h)
<U><U>
measure units “Ib”, “oz”, “kg”, “g”
<W><W><W><W><W><W>
weight data 6 Bytes
<H1><H2>
current status
Bit definition <H1 H2>
Bit Byte 1 (H1)
Byte 2 (H2)
0
0=stable
1=not stable
0=not under capacity
1=under capacity
1
0=not at zero point
1= at zero point
0=not over capacity
1=over capacity
2
0=RAM ok
1=RAM error
0=Flash ROM ok
1= Flash ROM error
3
0=eeprom ok
1=eeprom error
0=calibration ok
1= calibration error
4
always 1
always 1
5
always 1
always 1
6
always 0
always 0
7
parity
parity