4.2.6 Calibration Test
[
C.test
]
Calibration test procedures:
1. Long press the
Menu
button until you see
mM.E.N.U
.
2. Release the button and wait for the display to show
C.A.L
. Press the
Yes
button.
3. Short press the
No
button several times to navigate until you see
C.test
. Press the
Yes
button.
4.
0
and the calibration unit are flashing on the display based on the capacity and unit you set in the setup
menu. With no weight on the platform, press the
Yes
button to establish the zero point.
5. The display shows
--C--
while the zero point is recorded.
6. The calibration weight and the unit of the last time are flashing on the display. (For example,
015.000
kg).
6. To change the test calibration weight, you can do the steps below.
a) Short press the
No
button several times until the desired number appears.
NOTE
: press the
Back
button can decrease the digit.
b) Short press the
Yes
button to accept the number and move on to the next digit.
c) Repeat the process until all the digits are correct.
d) Press the
Yes
button to accept the calibration point. It is flashing on the display.
7. Place the specified test weight on the platform and press the
Yes
button.
8. The difference between the calibration data and the test weight is flashing on the display. (For example,
0.010
kg). If the terminal is connected to a printer or other devices, the result will be printed.
NOTE
: the printed result will appear as below:
Example:
--------- Cal Test ---------
New Cal:15.000kg
Old Cal:15.000kg
Diff Cal:0.0000kg
Wt. ID:_____________
------------ End -----------
9. After 5 seconds, the test ends and the scale returns to the active weighing mode, displaying the current
weight.
4.2.7 End Cal [End]
When
End
is displayed, you can press the
Yes
button to exit this menu and enter the next sub-menu, or press
the
No
button to enter to the first menu item in this sub-menu.
4.3 Setup Menu
You can set the scale parameters with the menu
S.E.t.u.P
. Default settings are in
bold
.
Menu Sub-Menu
Sub-Menu (in
segment)
Options
Options (in segment)
Setup
S.E.t.u.P
Reset
reset
no
, yes
NO, yes
Capacity Unit
C.UNIt
kg
, lb
/
Capacity
CAP
1-99999.9
/
Graduation
Grad
0.0001~100
/
Power On Zero
P.ZerO
Off,
On
Off
,
ON
Power On Unit
P.UNIt
Auto
, g, kg, lb, oz, lb:oz
AUtO
Auto Tare
A.tare
Off
, On
Off
,
ON
End
End
/
/
NOTE
: / indicates that it is not applicable in the table above.
Reset
[
reset
]
Reset the Setup menu to factory defaults.
NO
= do not reset
yes
= reset
Capacity Unit [
C.
UNIt
]
Select the unit used for calibration.
kg
lb