Courier 7000
EN-37
Request scale to tare
Command: T<CR> (54h,0dh)
simulate TARE key
<LF> <H1> <H2> <CR> <ETX>
Change units
of measure
Command:
U<CR> (55h,0dh)
simulate UNIT key
(Ib/kg)
<LF>
<U>
<U>
<CR>
<LF>
<H1>
<H2>
<CR>
<ETX>
simulate UNIT key
(Ib:oz)
<LF>
l
b
:
o
z
<CR>
<LF>
<H1>
<H2>
<CR>
<ETX>
Power off the scale
Command: X<CR> (58h,0dh)
simulate OFF key
x
x
x
x
x
Unrecognized Command
Command: others (xxh, 0dh)
Response
<LF>
?
<CR>
<ETX>
Definition of the Symbols and the Bits
Symbols Used
<LF>
line feed (0Ah)
<CR>
carriage return (0Dh)
<ETX>
end of text (03h)
<sp>
space (20h)
<p>
polarity “-” or “ ” (2Dh or 20h)
<U><U>
measure units “Ib”, “oz”, “kg”, “g”
<W><W><W><W><W>
weight data 5~6 Bytes
<H1><H2>
current status
Bit definition <H1-H2>
Bit Byte 1 (H1)
Byte 2 (H2)
0
0=stable
1=no stable
0=not under capacity
1=under capacity
1
0=not at zero point
1= at zero point
0=not over capacity
1=over capacity
2
0=RAM ok
1=RAM error
0=Flash ROM ok
1= Flash ROM error
3
0=eeprom ok
1=eeprom error
0=calibration ok
1= calibration error
4
always 1
always 1
5
always 1
always 1
6
always 0
always 0
7
parity
parity
6.3.2 8213 Interface Protocol
8213 Protocol
Request displayed weight
Command: W (57h)
over capacity
(invalid data)
<STX>
?
<S> <CR>
under
capacity
(-20d)
<STX>
?
<S> <CR>
under zero
(Mulis)
<STX>
?
<S> <CR>
not stable
<STX>
?
<S> <CR>
in Ib/oz/kg/g
(normal data)
<STX>
<W>
.
<W>
<W> <W> <W> <U> <U> <CR>
in Ib/oz/kg/g
<STX>
<W> <W>
.
<W> <W> <W> <U> <U> <CR>
Содержание Courier 7000
Страница 1: ...Courier 7000 Instruction Manual...
Страница 2: ......
Страница 4: ......
Страница 61: ...Courier 7000 EN 57 11 5 Column Kit Installation 1 2 3 4 5 6...