background image

 
 

 

EN-7

Catapult 1000 Series Scales

With the platform empty, press the 

ON/ZERO 

Off

 button to 

capture the scale zero. The display will show “-C-”.
 
 
After the zero is captured, the required span calibration 
weight in the selected unit is displayed. 
 
Place the calibration weight on the platform and press 

ON/ZERO 

Off

 button.  The display will again show “-C-”. 

 

After span capture, the display will return to the normal 
weighing mode. 

 

 
 

 

 
 

 

 

 
NOTE:

 The message “CAL E” will appear if an incorrect calibration weight was 

applied. Repeat the procedure using the correct calibration weight.

 

 

4.2 Calibration Locking 

 
By use of an internal switch, the calibration function may be locked (disabled), and 
unauthorized recalibration prevented. 
 
With the scale powered off, remove the cover under the indicator by unscrewing the 
two screws at the bottom.  

 

Slide the switch on the PCB (marked SW3) to the position marked “CAL LOCK”. 
 
Reassemble the cover.
 
If required, place tamper evident sealing labels over the screw holes or over the 
housing edges. 

 

If later recalibration is required, the cover will need to be reopened and the lock 
switch returned to the original position. 

 

NOTE: 

When the lock switch is set to the CAL LOCK position, the power-on zero 

range is reduced to 10% of full capacity. 

 

Содержание Catapult 1000 Series

Страница 1: ...00 Series TM Instruction Manual Serie Catapult 1000 TM Manual de instrucciones Catapult s rie 1000 TM Manuel d instructions Catapult 1000 Serie TM Bedienungsanleitung Serie Catapult 1000 TM Manuale di...

Страница 2: ......

Страница 3: ...Please follow these safety precautions Verify that the AC Adapter input voltage matches the local AC power supply Do not immerse the scale in water or other liquids Do not operate the scale in hostile...

Страница 4: ...d decimal places Prompts for calibration and possible error conditions are also shown using these digits 1 2 2 Stable Reading Indication A will appear in the lower left corner of the display to indica...

Страница 5: ...eady surface Avoid locations with excessive air current vibrations heat sources or rapid temperature changes 2 3 Power 2 3 1 AC Power The AC adapter included may be used to power the scale when batter...

Страница 6: ...accordance with local laws and regulations 3 OPERATION 3 1 Turning Scale On Off To turn the scale on press the ON ZERO Off button The scale performs a display test momentarily displays the model soft...

Страница 7: ...s of inactivity This feature is only active during battery operation 3 5 Dynamic Weighing With the platform empty short press the UNITS Cal button to cycle through the units Dynamic mode is active whe...

Страница 8: ...during the entire calibration process Calibration is unavailable in Dynamic Weighing Mode Allow the scale to warm up for at least 2 minutes after stabilizing to room temperature To abort calibration p...

Страница 9: ...using the correct calibration weight 4 2 Calibration Locking By use of an internal switch the calibration function may be locked disabled and unauthorized recalibration prevented With the scale powere...

Страница 10: ...ttery power used up Verify connections and voltage Verify battery polarity Poor accuracy Improper calibration Unstable environment Perform calibration Move scale to suitable location Unable to calibra...

Страница 11: ...E 6 1 SPECIFICATIONS Model C11P9 C11P20 C11P75 Capacity x Readability 9 kg x 0 005 kg 20 lb x 0 01 lb 20 kg x 0 01 kg 44 lb x 0 02 lb 75 kg x 0 05 kg 165 lb x 0 1 lb Weighing Units kg lb Modes Simple...

Страница 12: ...EN 10 Catapult 1000 Series Scales 6 2 Drawings and Dimensions Figure 6 1 Catapult 1000 Series Overall Dimensions...

Страница 13: ...Waste Electrical and Electronic Equipment WEEE this device may not be disposed of in domestic waste This also applies to countries outside the EU per their specific requirements Please dispose of this...

Страница 14: ...ent system is compliant with the ISO 9001 standard s requirements On May 15 2003 Ohaus Corporation USA was re registered to the ISO 9001 2000 standard LIMITED WARRANTY Ohaus products are warranted aga...