Catapult 1000 Series Scales

4. CALIBRATION 

For best results, calibrate the scale at regular intervals.  (Calibration weights are not 
supplied with the scale.) 

 

CAUTION 

USE EXTREME CARE WHEN HANDLING CALIBRATION WEIGHTS 
AS THEY ARE VERY HEAVY.  IMPROPER LIFTING METHODS OR 
MISUSE OF CALIBRATION WEIGHTS MAY RESULT IN PERSONAL 
INJURY.  MULTIPLE WEIGHTS MAY BE USED TO EQUAL THE 
REQUIRED CALIBRATION WEIGHT 

 

NOTE: 

•  Ensure the appropriate calibration masses are available before beginning 
    calibration (see Table 4-1).

 

 

•  Ensure that the scale is level and stable during the entire calibration 
    process.   
•  Calibration is unavailable in Dynamic Weighing Mode. 
•  Allow the scale to warm up for at least 2 minutes after stabilizing 
    to room temperature. 
•  To abort calibration, press 

UNITS 

Cal

, or power off the scale. 

 

TABLE 4-1. CALIBRATION WEIGHTS. 

Model 

Calibration weight (kg/lb) 

C11P9

9 / 20

C11P20

20 / 40

C11P75

50 / 100

 

4.1 Span Calibration

Span Calibration uses two points to adjust the scale.  The first point is the zero value 
where there is no weight on the scale.  The second point is the Span value where a 
calibration mass is placed on the scale. 
 
To select the weighing unit to be used for calibration, 
press the 

UNITS 

Cal

 button until the correct unit is 

displayed.  

 

Press and hold the 

UNITS 

Cal

 button until “CAL” is 

displayed. 

 
 
 
 

 

 

Содержание Catapult 1000 Series

Страница 1: ...00 Series TM Instruction Manual Serie Catapult 1000 TM Manual de instrucciones Catapult s rie 1000 TM Manuel d instructions Catapult 1000 Serie TM Bedienungsanleitung Serie Catapult 1000 TM Manuale di...

Страница 2: ......

Страница 3: ...Please follow these safety precautions Verify that the AC Adapter input voltage matches the local AC power supply Do not immerse the scale in water or other liquids Do not operate the scale in hostile...

Страница 4: ...d decimal places Prompts for calibration and possible error conditions are also shown using these digits 1 2 2 Stable Reading Indication A will appear in the lower left corner of the display to indica...

Страница 5: ...eady surface Avoid locations with excessive air current vibrations heat sources or rapid temperature changes 2 3 Power 2 3 1 AC Power The AC adapter included may be used to power the scale when batter...

Страница 6: ...accordance with local laws and regulations 3 OPERATION 3 1 Turning Scale On Off To turn the scale on press the ON ZERO Off button The scale performs a display test momentarily displays the model soft...

Страница 7: ...s of inactivity This feature is only active during battery operation 3 5 Dynamic Weighing With the platform empty short press the UNITS Cal button to cycle through the units Dynamic mode is active whe...

Страница 8: ...during the entire calibration process Calibration is unavailable in Dynamic Weighing Mode Allow the scale to warm up for at least 2 minutes after stabilizing to room temperature To abort calibration p...

Страница 9: ...using the correct calibration weight 4 2 Calibration Locking By use of an internal switch the calibration function may be locked disabled and unauthorized recalibration prevented With the scale powere...

Страница 10: ...ttery power used up Verify connections and voltage Verify battery polarity Poor accuracy Improper calibration Unstable environment Perform calibration Move scale to suitable location Unable to calibra...

Страница 11: ...E 6 1 SPECIFICATIONS Model C11P9 C11P20 C11P75 Capacity x Readability 9 kg x 0 005 kg 20 lb x 0 01 lb 20 kg x 0 01 kg 44 lb x 0 02 lb 75 kg x 0 05 kg 165 lb x 0 1 lb Weighing Units kg lb Modes Simple...

Страница 12: ...EN 10 Catapult 1000 Series Scales 6 2 Drawings and Dimensions Figure 6 1 Catapult 1000 Series Overall Dimensions...

Страница 13: ...Waste Electrical and Electronic Equipment WEEE this device may not be disposed of in domestic waste This also applies to countries outside the EU per their specific requirements Please dispose of this...

Страница 14: ...ent system is compliant with the ISO 9001 standard s requirements On May 15 2003 Ohaus Corporation USA was re registered to the ISO 9001 2000 standard LIMITED WARRANTY Ohaus products are warranted aga...

Отзывы: