Catapult 1000 Series Scales

5. MAINTENANCE 

Caution: before cleaning, turn off the scale, and remove the AC adapter. 
 

5.1 Cleaning 

The housing may be cleaned with a cloth dampened with a mild detergent if 
necessary. 
Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the 
housing or control panels. 
 

5.2 Troubleshooting 

The following table lists common problems and possible causes and remedies.  If 
the problem persists, contact Ohaus or your authorized dealer. 
 
                                  

TABLE 5.1. TROUBLESHOOTING.

Symptom Possible 

Cause(s) 

Remedy 

Scale will not turn on   No power to scale 

Battery power used up 

Verify connections and voltage. 
Verify battery polarity 

Poor accuracy 

Improper calibration. Unstable 
environment  

Perform calibration. Move scale to 
suitable location  

Unable to calibrate 

CAL LOCK set to ON 
Unstable environment 
Incorrect calibration mass 

Set CAL LOCK to OFF (see section 
4.2) 
Move the scale to suitable location 
Use correct calibration mass 

Scale displays “Lo 
bAt” 

Battery is discharged 

Connect power or change the 
batteries 

Scale displays “Lo 
Line” 

Low Line voltage 

Verify source voltage 

Scale displays “E” 

Weight on pan exceeds 
capacity  

Remove weight from the pan  

Scale displays “CALE”  Incorrect Calbiration weight on 

pan 

Use correct calibration weight  

Scale displays “UnSt”  Scale is unstable at power up.  Relocate scale to a more stable 

environment 

 

Содержание Catapult 1000 Series

Страница 1: ...00 Series TM Instruction Manual Serie Catapult 1000 TM Manual de instrucciones Catapult s rie 1000 TM Manuel d instructions Catapult 1000 Serie TM Bedienungsanleitung Serie Catapult 1000 TM Manuale di...

Страница 2: ......

Страница 3: ...Please follow these safety precautions Verify that the AC Adapter input voltage matches the local AC power supply Do not immerse the scale in water or other liquids Do not operate the scale in hostile...

Страница 4: ...d decimal places Prompts for calibration and possible error conditions are also shown using these digits 1 2 2 Stable Reading Indication A will appear in the lower left corner of the display to indica...

Страница 5: ...eady surface Avoid locations with excessive air current vibrations heat sources or rapid temperature changes 2 3 Power 2 3 1 AC Power The AC adapter included may be used to power the scale when batter...

Страница 6: ...accordance with local laws and regulations 3 OPERATION 3 1 Turning Scale On Off To turn the scale on press the ON ZERO Off button The scale performs a display test momentarily displays the model soft...

Страница 7: ...s of inactivity This feature is only active during battery operation 3 5 Dynamic Weighing With the platform empty short press the UNITS Cal button to cycle through the units Dynamic mode is active whe...

Страница 8: ...during the entire calibration process Calibration is unavailable in Dynamic Weighing Mode Allow the scale to warm up for at least 2 minutes after stabilizing to room temperature To abort calibration p...

Страница 9: ...using the correct calibration weight 4 2 Calibration Locking By use of an internal switch the calibration function may be locked disabled and unauthorized recalibration prevented With the scale powere...

Страница 10: ...ttery power used up Verify connections and voltage Verify battery polarity Poor accuracy Improper calibration Unstable environment Perform calibration Move scale to suitable location Unable to calibra...

Страница 11: ...E 6 1 SPECIFICATIONS Model C11P9 C11P20 C11P75 Capacity x Readability 9 kg x 0 005 kg 20 lb x 0 01 lb 20 kg x 0 01 kg 44 lb x 0 02 lb 75 kg x 0 05 kg 165 lb x 0 1 lb Weighing Units kg lb Modes Simple...

Страница 12: ...EN 10 Catapult 1000 Series Scales 6 2 Drawings and Dimensions Figure 6 1 Catapult 1000 Series Overall Dimensions...

Страница 13: ...Waste Electrical and Electronic Equipment WEEE this device may not be disposed of in domestic waste This also applies to countries outside the EU per their specific requirements Please dispose of this...

Страница 14: ...ent system is compliant with the ISO 9001 standard s requirements On May 15 2003 Ohaus Corporation USA was re registered to the ISO 9001 2000 standard LIMITED WARRANTY Ohaus products are warranted aga...

Отзывы: