Страница 1: ...Adventurer Pro Grain Test Scale Instruction Manual ...

Страница 2: ...se harmful interference in which case the user will be required to correct the interference at his own expense This Class A digital apparatus complies with Canadian ICES 003 Cet appareil numérique de la classe A est conforme à la norme NMB 003 du Canada ISO 9001 Registration In 1994 Ohaus Corporation USA was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International ...

Страница 3: ...rview of Display Functions EN 8 3 3 Button Control Functions EN 9 3 4 Using the Button Control Functions EN 10 3 4 1 Setting the Scale to Zero EN 10 3 4 2 Taring EN 10 3 4 3 Changing Units of Measure EN 10 3 4 4 Printing Data EN 10 3 5 Menu EN 11 3 5 1 Menu Structure EN 11 3 5 2 Menu Navigation EN 12 3 5 3 Changing Settings EN 12 3 6 Application Modes EN 13 3 6 1 Weighing EN 13 3 6 2 Test Weight M...

Страница 4: ...Sealing the Scale EN 24 3 9 Printing Data EN 25 4 MAINTENANCE EN 27 4 1 Calibration EN 27 4 2 Cleaning EN 27 4 3 Troubleshooting EN 27 4 4 Service Information EN 28 4 5 Parts EN 28 4 6 Accessories EN 29 5 TECHNICAL DATA EN 31 5 1 Outline Drawing EN 31 5 2 Specifications EN 32 5 3 Capacity x Readability Table EN 32 5 4 Communication EN 33 5 4 1 Commands EN 33 5 4 2 Connections EN 34 Limited Warrant...

Страница 5: ...office for further details 1 1 Description The Adventurer Pro Grain Test Scale is a precision weighing instrument that will provide you with years of service if properly cared for The Adventurer Pro Grain Test Scale has a capacity of 4100 grams 1 2 Features The Adventurer Pro Grain Test Scale includes many standard features They are AC Adapter operation AC Adapter included Weighing Test Weight Doc...

Страница 6: ...Components Use the illustration and instructions below to identify and assemble your Adventurer Pro Grain Test Scale with its components All components must be installed before using the scale 2 2 1 Assembly 1 Insert the Sub platform on the Mounting Cone located in the center of the scale Align the Sub platform so that it is fully seated on the cone 2 Place the Weighing Pan on the Sub platform Sub...

Страница 7: ...bration procedure 2 3 Selecting the Location Locate the scale on a firm steady surface Avoid locations with excessive air current vibrations heat sources or rapid temperature changes 2 4 Leveling the Scale Before the scale is used the feet should be adjusted so that the scale is level with the surface This will enable accurate weighing You will see a level bubble indicator in a small round window ...

Страница 8: ...Adventurer Pro Grain Test Scale EN 6 ...

Страница 9: ...1 ADVENTURER PRO GRAIN TEST SCALE CONTROLS No Description 1 ON O T Off Yes Button 2 UNIT Print No Button 3 Level Bubble 4 Recall Back button 5 TEST WEIGHT Menu Cal Exit Button 6 Display 7 Pan 8 Type label on side of housing 9 Data label 10 Leveling feet 11 Power Connector 12 COM 2 Connector AV R and AV U Models only 13 Security Loop 14 Lock Switch 15 COM 1 Connector 9 ...

Страница 10: ...iption 1 Primary 7 segment display 2 Brackets 3 Units of Measurement 4 Memory Indicator 5 Battery Strength Indicator not used 6 Pieces Indicator not used 7 Secondary 14 Segment Display 8 Gross Indicator not used 9 Brutto Indicator not used 10 Tare Indicator not used 11 Preset Tare Indicator not used 12 Net Indicator not used 13 Stability Indicator 14 Negative Indicator 15 Center of Zero Indicator ...

Страница 11: ...utton is shown below Press and Release Menu Function On O T If scale is off turns scale on If scale is on sets zero Off Turns scale off Yes Accepts the current blinking setting on the display UNIT Change weighing unit Print Sends current display value to the serial interface No Rejects the current setting blinking on the display Increments a value being entered TEST WEIGHT Activates selected Test ...

Страница 12: ...e Tare value remove the container from the pan and press the ON O T button Preset Tare Use the xT command in the command table to enter a preset tare value through a computer To clear the Tare value enter a value of 0 0 See Section 5 4 1 3 4 3 Changing Units of Measure The Adventurer Pro Grain Test Scale can be configured to measure in a variety of units including custom units The Unit Sub menu is...

Страница 13: ...en 7Odd 7 No par 8 No par Handshake Off XONXOFF End RS1 RS232 2 Optional Baud 600 2400 19 200 Parity 7Even 7Odd 7 No par 8 No par Handshake Off XONXOFF End RS1 GLP DATA 3 8 9 User ID Set Proj ID Set Time Type Set Adj Date Type Set End GLP Data GLP PRINT 3 8 10 Time On Off Balance ID On Off User ID On Off Project ID On Off Difference On Off Name On Off End GLP Prt PRINT 2 Optional Output When Stabl...

Страница 14: ...d the Menu button until MENU appears on the primary 7 Segment display Release the button and the first sub menu appears on the secondary 14 Segment display blinking Note When the secondary display is blinking press Yes to accept Press No or Back or view another selection SELECT THE SUB MENU Press the No button to select the next sub menu or the Back button to select the previous sub menu in the lo...

Страница 15: ...Weight Mode The Test Weight Mode allows the measurement of the density of a sample using one of four standard container sizes Test Weight values can be displayed as follows lb bu qt pounds per bushel based on a quart container lb bu pt pounds per bushel based on a pint container kg hL L kilograms per hectaliter based on a liter container kg hL SL kilograms per hectaliter based on a half liter cont...

Страница 16: ...s not accepted as shown If the sample weight is greater than or equal to 10 grams the sample weight is established as the Dockage Reference weight and is stored in memory as the 100 reference weight for the dockage calculation Additionally the tare weight is stored in memory as the dockage tare weight With the sample still on the pan the primary display shows 100 00 if Positive Dockage is selected...

Страница 17: ...weight the container with the processed sample is placed on the pan and the reference weight is recalled from memory by a long press of the Recall button The primary display shows the percent dockage of the sample based on the previously established dockage reference weight and dockage tare weight Percent Dockage Display The percent dockage display capacity is fixed at 410 09 The percent dockage d...

Страница 18: ...earity Linearity calibration minimizes deviation between actual and displayed weights within the scale s weighing range Three weight values are used zero a weight value at midpoint of the scale s weighing range and a weight value at or near the scale s specified capacity Cal Test Cal Test Cal Test Cal Test Cal Test Calibration test allows the stored calibration data to be tested against the curren...

Страница 19: ...when the desired calibration weight is displayed Place the specified calibration weight on the pan After calibration remove the weight from the pan To abort calibration at any time press the Exit button Span Calibration NOTICE Linearity Calibration is required only if the linearity error exceeds tolerance in specification table Linearity Calibration With the scale on press and hold the Menu Cal bu...

Страница 20: ... 7 2 Calibration Masses Model Linearity Span Weight Class Calibration Points Calibration Points ASTM OIML AVD4102N 2000g 4000g 2000g 3000g 4000g 1 E2 Calibration Test Calibration test allows a check of a known calibration mass against the last stored calibration information in the scale Press and hold the Menu Cal button until MENU CALIBRATE is displayed Press Yes to enter calibration Press No unt...

Страница 21: ...w Medium High Compensates for vibration or excessive air currents READOUT Auto Zero Off 5d 1d 2d 5d Filter Low Medium High End Read 3 7 4 Readout 3 7 5 Mode The Mode submenu is used to set Application modes parameters MODE Weigh On Off Test Weight lb bu qt lb bu pt kg hL L kg hL 5L Dockage Pos Neg End Mode Weigh Use to set the Weighing Mode On or Off Test Weight Use to set the Test Weight to lb bu...

Страница 22: ...ional COM2 port is installed Enter the Factor following the instructions for Editing Limits in Section 3 5 4 Enter the Exponent and LSD using the Yes No buttons Factor Exponent Conversion 3 to 3 Factor 1234 3 123 4 1234 2 12 34 1234 1 1 234 1234 0 1234 1234 1 01234 1234 2 001234 1234 3 0001234 PRINT 1 Output When Stable On Off GLP Tare On Off Auto Print Off Cont Interval When Stable 1 3600 seconds...

Страница 23: ...ers for an external printer or computer RS232 2 contains an identical menu and is available when the optional COM2 port is installed RS232 1 Baud 600 2400 19200 Parity 7 Even 7 Odd 7 No Par 8 No Parity Handshake Off XONXOFF Hardware End RS1 Content All of these features can be set On or Off Numeric data only Header Weight Reference Result GLP See sample printout 3 9 GLP DATA User ID Set Proj ID Se...

Страница 24: ...ustment value from 60 to 60 seconds per day Date Set the date type M D Y D M Y Y M D M Y D Y D M D Y M and the actual date End GLP Data Press the Yes button to advance to the GLP Print submenu Press the No button to return to User ID 3 7 10 GLP Print Select GLP items to be printed by setting them to On GLP PRINT Time On Off Balance ID On Off User ID On Off Project ID On Off Difference On Off Name ...

Страница 25: ...l Select Yes to return the menu items in all sub menus to their factory settings End Reset Press the Yes button to advance to the Lockout menu Press the No button to return to Setup 3 7 12 Lockout Use this submenu to prevent unauthorized changes to menu settings If a submenu is locked its menu item settings can be viewed but not changed LOCKOUT Cal On Off Setup On Off Readout On Off Mode On Off Un...

Страница 26: ...ked In the Print Sub menu the Continuous setting in the Auto Print menu item is hidden and not available In the Reset Sub menu the Reset Readout Reset Mode Reset Unit and Reset Global menu items are locked In the Lockout Sub menu the Calibration menu item is locked To regain access to the locked Menu settings turn the scale off Turn the scale on and press the Lock Switch while OHAUS is displayed d...

Страница 27: ...ne 4 If none of the 5 header lines are entered Line 5 using the H x command no header is printed If Print x Content GLP is ON 04 01 06 12 30 00 PM If GLP PRINT Time is ON Bal ID 1234567 If GLP PRINT Balance ID is ON USER ID ABCDEFGHIJ If GLP PRINT User ID is ON PROJ ID 1234567890 If GLP PRINT Proj ID is ON Name If GLP PRINT Name is ON 0 0200kg Weight is printed if selected in the definition string...

Страница 28: ...e 5 using the H x command no header is printed LIN CAL 04 01 06 12 30 00 PM If GLP PRINT Time is ON Bal ID 1234567 If GLP PRINT Balance ID is ON Cal 4000 00g Old Cal 3999 96g Add Mid Point and Zero Diff 0 04g Wt ID_______________ USER ID ABCDEFGHIJ iF GLP PRINT User ID is ON PROJ ID 1234567890 If GLP PRINT User ID is ON Name If GLP PRINT Name is ON END OHAUS Header Line 1 Line 2 Header Line 2 Line...

Страница 29: ... Weight reading exceeds Clear pan Power On Zero limit Error 8 2 Weight reading below Put pan on scale Power On Zero limit Error 8 3 Weight reading exceeds Clear pan overload limit Error 8 4 Weight reading below Put pan on underload limit scale Error 9 0 Internal fault Return scale for service 4 MAINTENANCE 4 1 Calibration Periodically verify calibration by placing an accurate weight on the scale a...

Страница 30: ... on the pan is too small to define a valid reference weight Busy tare zero printing Wait until completion 4 3 Troubleshooting Cont 4 4 Service Information If the Troubleshooting section does not resolve or describe your problem contact your authorized Ohaus Service Agent For Service assistance or technical support in the United States call toll free 1 800 526 0659 between 8 00 AM and 5 00 PM EST A...

Страница 31: ...vice 76288 01 RS232 Cables Cable DB9M DB9F 80500525 Cable DB9M DB25F 80500524 Cable DB9M SF42 Printer 80500571 Cable DB9M Apple 80500562 SF42 Printer SF42 4 5 Parts Cont Description Ohaus Part Number Replacement Pan Rectangular pan 7 6 x 8 in 19 3 x 20 3 cm 12103880 ...

Страница 32: ...Adventurer Pro Grain Test Scale EN 30 ...

Страница 33: ...t ambient temperatures between 5 and 40 C Power AC adapter To national standard as per list in Section 4 6 Scale power input 6 14 5 VAC 50 60Hz 4VA or 7 20VDC 4W Materials Housing base plastic ABS PC Top Housing plastic ABS PC Platform 18 10 stainless steel Protection Protected against dust and water Pollution degree 2 Installation category Class II EMC see declaration of conformity Overall Dimens...

Страница 34: ...ternal Weight Display Type 2 Line LCD w Backlight Display Size in cm 4 x 1 10 x 2 5 Platform Size in cm 6 6 x 7 1 16 8 x 18 Dimensions 8 7 x 3 3 x 11 8 W x H x D in mm 22 x 8 5 x 30 Net Weight lb kg 6 1 2 8 5 2 Specifications 5 TECHNICAL DATA 5 3 Capacity x Readability Table Unit LPT OFF COMMERCIAL GIPSA grain 63272 8 x 0 2 63272 x 2 63272 x 2 gram 4100 00 x 0 01 4100 0 x 0 1 4100 0 0 x 0 0 1 kilo...

Страница 35: ...nstable P Print displayed weight stable or unstable CP Continuous Print 0P ends Continous Print SP Print displayed stable weight SLP Auto Print stable non zero displayed weight SLZP Auto Print stable non zero weight and stable zero reading xP Interval Print x Print Interval 1 3600 seconds 0P ends Interval Print H Enter Print Header Lines Z Same as pressing Zero Key T Same as pressing Tare Key PM P...

Страница 36: ...S 9 Remote Print COM 2 MINI DIN Connector TARE PRINT External PRINT and or TARE switches may be installed as shown in the diagram Momentary contact switches must be used To enable this feature contact Ohaus Aftermarket A 6 pin Mini DIN connector is provided when the optional second RS232 interface is installed 1 2 3 4 5 6 7 8 9 COM2 Pin Connections 1 TxD 2 RxD 3 Ground 4 Vout 5 reserved for future...

Страница 37: ...duct has been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than Ohaus In lieu of a properly returned warranty registration card the warranty period shall begin on the date of shipment to the authorized dealer No other express or implied warranty is given b...

Страница 38: ...ration 19A Chapin Road P O Box 2033 Pine Brook NJ 07058 USA Tel 973 377 9000 Fax 973 593 0359 80103489 With offices worldwide www ohaus com Ohaus Corporation 2006 all rights reserved Printed in China P N 80103489 ...