EN 40
7000 SERIES INDICATORS
The use of Alibi Memory requires installation of an SD memory card. Attempts to generate a print transaction from
the front panel without an SD card installed return a pop-up error message SD card is not installed., while a print
command issued serially or via an input will result in an SD memory card not installed message in the system line.
The transaction counter will not increment.
The Alibi Memory table stores basic transaction information that is not user-definable. This information always
includes:
Date and time stamp
Transaction counter value
Gross, tare, and net weights and weight unit
If the T72XW terminal has been programmed as “approved”, Alibi Memory enabling or disabling is only
accessible if the security switch (SW1-1) is in the OFF position.
Totalization
Use the Totalization setup screen to select parameters for totalization operations, including which source to use as
input for totalization, settings for grand totals and subtotals, and to enable or disable the conversion of second unit
weights for totalization. When a demand print occurs, the selected weight field will be added to the totals register.
Mode
Select which source to use as input for totalization comparison
– either Displayed Weight or Gross Weight. A
selection of None disables totalization.
Clear GT on Print.
The Grand Total can be configured to clear automatically after printing the Totals report. If Clear GT on Print is
Enabled, the subtotal also clears automatically after printing the Totals report.
Subtotal
The Subtotal can be separately disabled while the GT continues to accumulate weights. Choose to either Enable or
Disable the subtotal register.
Clear ST on Print
Clearing the subtotal on print and not clearing the grand total on print allows the subtotal register to totalize sub-
sets of weighments and to be reset while the grand total continues to track the grand total of weight. Choose to
Clear ST on Print or not by selecting Enabled or Disabled in the selection box..
Convert Weight
The total registers always store weights in primary units. If Convert Weight is Disabled, scale weights other than
primary units are not accumulated. If Convert Weight is Enabled, then the weight is converted to primary units, then
accumulated.
Discrete I/O
Discrete I/O setup menus allow the configuration of 2 inputs and 4 outputs. The Discrete I/O option must be
installed to program this functionality.
Inputs
The Discrete Input menus display discrete input polarity and assignments for Input 1 and Input 2. The two Inputs
are configured in the same way.
Polarity
The inputs can be programmed to accept either a + True or
– True polarity level as “ON”. The default is + True.