7000 SERIES INDICATORS
EN 103
I1
– INQUIRY OF MT-SICS LEVEL AND MT-SICS VERSIONS
Command:
I1
Response:
I1_A_“ ”_“2.2x”_“2.2x”_“ ”_“ ”
“” No Levels fully implemented
2.2x
Level 0, version V2.2x
2.2x
Level 1, version V2.2x
“” No MT-SICS 2 commands
“” No MT-SICS 3 commands
Response:
I1_I
Command understood, not executable at present.
Comments
For the MT-SICS level, only fully implemented levels are listed. In this case, neither level 0 nor level 1 were
fully implemented so the level is not specified.
In the case of the MT-SICS version, all levels are specified even those only partially implemented.
I2
– INQUIRY OF DATA
Command:
I2
Response:
I2_A_“T72XW_Standard_50.00_kg”
T72XW
Model number of terminal
Standard
Standard functionality
50.00 kg
Capacity and primary unit of the scale
Response:
I2_I
Command understood, not executable at present.
Comments
The number of characters in the “text” field depends on the scale capacity.
I3
– INQUIRY OF SW VERSION
Command:
I3
Response:
I3_A_“1.00”
1.00
Firmware version of the terminal
Response:
I3_I
Command understood, not executable at present.
Comment
The number of characters of “text” depends on the revision level.
I4
– INQUIRY OF SERIAL NUMBER
Command:
I4
Response:
I4_A_“123456”
123456 Serial number of terminal
Response:
I4_I
Command understood, not executable at present.