EN-3
3000 Series Indicators
TABLE -. TROUBLESHOOTING (Cont.).
SYMPTOM
PROBABLE CAUSE(s)
REMEDY
Error 8.1
Weght readng exceeds Power On Zero lmt.
Remove load from scale. Recalbrate scale.
Error 8.2
Weght readng below Power On Zero lmt.
Add load to scale. Recalbrate scale.
Error 8.3
Weght readng exceeds Overload lmt.
Reduce load on scale.
Error 8.4
Weght readng below Underload lmt.
Add load to scale. Recalbrate scale.
Err 9.0
Internal fault
Servce requred.
Err 9.5
Calbraton data not present.
Calbrate scale.
Err 53
EEPROM data ncorrect.
Servce requred.
CAL E
Calbraton Error. Calbraton value outsde
allowable lmts.
Repeat calbraton usng correct calbraton
weghts.
LOW.rEF
The average pece weght of the parts s small
(warnng).
Use parts wth average pece weght greater
than or equal to 1 dvson.
REF.WT Err
The average pece weght of the parts s too
small.
Use parts wth a average pece weght
greater than or equal to 0.1 dvson.
.3 Service Information
If the troubleshootng secton does not resolve your problem, contact an authorzed Ohaus Servce Agent. For Servce assstance
n the Unted States, call toll-free 1-800-526-0659 between 8:00 AM and 5:00 PM Eastern Standard Tme. An Ohaus Product
Servce Specalst wll be avalable to assst you. Outsde the USA, please vst our webste www.ohaus.com to locate the Ohaus
offce nearest you.