OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
41
Rotor Assembly
Rotor Assembly with
Slurry Cup Table
7. Thread the slurry cup table onto the rotor shaft assembly.
Pour a small amount of mineral oil into the vessel. This will act as a
cushion when inserting the rotor assembly.
8. Insert the rotor assembly into the drive housing. Press down on the
slurry cup table until it falls into place.
9. Replace the cover and gland underneath the test cell before beginning
another test.
Filters
The instrument has an air filter and an air dryer. Both are located inside
the cabinet on the left-hand side (see photo on page 44). Periodically
check both of these for accumulated water. Unscrew the plug on the
bottom and let the water drain.
The instrument has two oil filters. Both are inside the cabinet. The high-
pressure filter is on the right-hand side and the low-pressure filter is on
the bottom below the cell (see photo on page 44). The low-pressure
filter should be replaced yearly. The high-pressure filter should be cleaned
when the flow of oil back into the reservoir is obstructed (see page
45 for instructions).