14
Status Indicator:
The lights on the Status Indicator correspond to the consistency of the
sample. The threshold values can be set in either the onboard controls or
in the software. There are three setpoints, specified in Bearden units of
consistency (Bc).
- Max Bc for Green Light: When the consistency is below this value, the
green light will be on.
- Max Bc for Green and Yellow: When the consistency is below this
value, the green and yellow lights will both be on.
- Max Bc for Red Light: When the consistency is below this value, the
yellow light only will be on. When the consistency is above this value,
the red light will be on and the test will stop.
Refer to page 18 for instructions on setting these values.
The Status Indicator also indicates an alarm condition. If an alarm is
triggered, all three lights on the status indicator will blink until the alarm is
acknowledged.
Switches:
-
FILL/DRAIN CELL: Controls the flow of oil into the cell.
-
DRAIN: Oil flows out of the cell and into the reservoir.
- OFF: The oil is not moved in either direction.
-
FILL: Oil flows out of the reservoir and into the cell for testing.
- POT VOLTS: Controls power to the potentiometer
- ON: Power is supplied to the potentiometer.
- AUTO: Power is only supplied to the potentiometer only during a
test or calibration.
- DAQ PAUSE:
- PAUSE: No test data is collected
- OPERATE: Test data is collected
- MOTOR: Sets control of the motor
- AUTO: The motor will turn according to the parameters of the test
that is currently running. If no test is running, the motor will be
stopped.
- OFF: The motor will not turn at all.
- ON: The motor will turn at 150 RPM until stopped.
-
ALARM: Turns the audible alarm on or off.