28
On the Test Config tab of the Test Setup screen, you can set thresholds for
the indicator lights on the top of the unit.
1. Go to the Test Setup screen and select “Test Config”.
2. Enter a consistency value (in Bc) for each threshold:
Signal LED G/Y:
When the consistency reaches this level, the green and
yellow lights will come on.
Signal LED Y:
When the consistency reaches this level, the yellow light
will come on.
Signal LED Y/R:
When the consistency reaches this level, the yellow and
red lights will come on.
Signal LED R:
When the consistency reaches this level, the red light will
come on.
3. Select “Save” when done.
Onboard
Display
Test Setup - Test
Config