4. Click OK to connect to the viscometer.
5. Press the Enter key twice to bring up the prompt.
6. Turn the bob counterclockwise and hold it against the stop.
7. On the computer type C and press Enter.
8. Ignore the instruction on the screen. Slowly turn the bob clockwise until it
reaches the other stop.
9. Type N and press Enter. The software will show details of the calibration.
Close Hyperterminal.