Installation
47
GA03107_002_A4 - 09/2015 - © Oerlikon Leybold Vacuum
Fig. 3.8 Checking the axial distance dimension for the coupling
L1
Fig. 3.9 View of the fitted motor
Washer (4x)
WA 251 / 501: Ø 11
WA 1001 / 2001: Ø 13
Nut (4x)
WA 251/501:
M10 / torque 40 Nm
WA 1001/2001: M12 / torque 40 Nm
L1
3.5.3 Fitting the Motor
At the pump’s flange the respective studs have been preassembled, the
washers and nuts which must be used have been enclosed with the pump.
Check to ensure that the touch protection (Fig. 3.6) has been correctly fitted
in the pump’s flange.
Insert the coupling star into the coupling half on the side of the pump.
Align the coupling halves and push the motor on making sure that the flange
surfaces run in parallel.
Any tilt between motor axis and impeller axis can give rise to increased wear
on the coupling star.
Fit washers and nuts, and tighten according to the torque levels given in Fig.
3.9.
3.5.4 Electrical Connection
Connect the motor according to Section 3.3 and to the information of the
manufacturer of the motor.
L1 = Axial distance dimension for the coupling
WA 251
WA 501
WA 1001
WA 2001
L1 (± 0.2 mm)
54.3 mm
63.3 mm
74.6 mm
100.6 mm
Set screw M8 / torque 10 Nm
Содержание RUVAC WA
Страница 67: ...Notes 67 GA03107_002_A4 09 2015 Oerlikon Leybold Vacuum ...
Страница 68: ...68 GA03107_002_A4 09 2015 Oerlikon Leybold Vacuum EC Declaration of Conformity ...
Страница 69: ...69 GA03107_002_A4 09 2015 Oerlikon Leybold Vacuum EC Incorporation Declaration ...
Страница 71: ...Index 71 GA03107_002_A4 09 2015 Oerlikon Leybold Vacuum ...