How must the Temperature Sensor be Connected Electrically?
Oerlikon Leybold Vacuum describes here a recommended circuit for the elec-
trical connection. The circuit shown (Fig. 0.3) represents one possibility of
processing the signal, other circuit implementations can produce the same
results!
Due to the installation position and the tolerances stated for the temperature
measurement, and for compliance with temperature classification TX, the
RUVAC WA/WAU must be shut down reliably at the shutdown temperatures
stated in Table 1. T4 operation is excluded.
24-60 V DC / AC or 85-230 V DC / AC
Combined measurement transducer
limit value indicator SONEAX VC 603
Analog output
Explosion
hazard area
Temperature sensor (PT 100)
0
5
4
4
1
6
11
Z
/26
/21
/21
/29
Fig. 0.3 Connection of the temperature sensor (circuit recommendation)