Interfaces
44
130000761_002_A0 - 08/2008 - © Oerlikon Leybold Vacuum
3.3.5 Parameter list
r = readable, w = writable
No. Designation
Min.
Max. Default Unit
r/w Format Description
1
Converter type
0
65535
0
r
u16
136 = Turbo.Drive 400
2
Software version
0
65535
10000
r
u32
xx.yy: version, zz: correction index
3
Actual frequency
0
65535
0
Hz
r
u16
Actual rotor frequency
4
Actual intermediate
0
1500
30
0,1 V
r
u16
Actual intermediate circuit voltage
circuit voltage
of the converter
5
Actual current
0
150
0
0,1 A
r
u16
Actual motor current
6
Actual electrical power
0
65535
0
0,1 W
r
u16
Actual drive input power
7
Actual motor
-10
150
0
°C
r
i16
Actual value of the motor
temperature
temperature.
8
Save data command
0
65535
0
/w
i16
A write command with any value saves
temporary data into nonvolatile memory.
11
Actual converter
-10
150
0
°C
r
i16
Actual heat sink temperature
temperature
of the converter.
16
Motor temperature
0
150
54
°C
r
i16
Exceeding the motor temperature
warning threshold
warning threshold results in a warning.
17
Nominal motor current
5
60
5,0
0,1 A
r
u16
Maximum permissible motor current
18
Maximum frequency
750
1200
1000
Hz
r
u16
Highest permissible frequency
19
Minimum frequency
0
1200
525
Hz
r
u16
Lowest permissible frequency
20
Critical frequency
0
1200
500
Hz
r
u16
Minimum frequency level. When the
pump is accelerating this frequency
must be reached within the
maximum passing time (P183).
23
Pump type
0
255
2
r
u16
6= SL 300
24
Setpoint frequency
0
1200
1000
Hz
r/w
u16
Setpoint of the rotor frequency
25
Normal operation
35
99
95
%
r/w
u16
Setpoint of the frequency dependent
normal operation level
27
Current norm. oper.
5
60
20
0,1 A
r/w
u16
Motor current dependent normal operation
level; ; If P29[0] = 1: Defines the normal
operation level. Normal operation if
P5 <= P27 Parameter cannot be changed
during operation of the system
29
Relay function X1
0
8
0
r/w
u16
If required, special functions can be
assigned to the normal operation
and the error relay.
Field 0 specifies the function for normal operation:
0 = Frequency dependent
1 = Motor current dependent
2 = Fieldbus controlled
3 = Trigger current bearing temperature (P122)
4 = Venting function (P247/P248)
5 = Pump at standstill (f < 3)
6 = Start command is present
7 = Ready for switch on (=STW Bit1)
8 = No mains power failure or no generator operation (P303 Bit 4 =1 = generator operation)
Field 1 specifies the function for the error relay:
0 = Energised when an error is present
1 = Deenergised when an error is present
2 = Fieldbus controlled